基板设计与骨髓脂肪浸润对胫骨基板拔出强度的生物力学比较
在第四代锯骨胫骨模型上比较PFC Sigma(对照)、Attune(第一代)和Attune S+(第二代)胫骨基板的拔出强度,每种基板各有三件在含或不含牛骨髓脂肪条件下骨水泥固定。第二代基板失效所需力显著更大,牛骨髓存在显著降低整体拔出力和所有设计,第一代与对照无显著差异。
要点
- 第二代胫骨基板失效所需力显著大于对照和第一代。
- 骨水泥固定时牛骨髓脂肪存在显著降低胫骨基板轴向拔出强度。
- 第一代基板均出现骨水泥-植入物界面脱粘。
核心结论
- 第二代基板失效所需力显著大于较老设计。
- 在该实验室模型中,骨水泥固定胫骨基板时骨髓存在显著降低轴向拔出强度。
- 所有第一代基板均出现骨水泥-植入物界面脱粘。
研究设计与样本
- 将PFC Sigma(对照)、Attune(第一代)和Attune S+(第二代)胫骨基板用标准化技术植入第四代锯骨胫骨模型。
- 每种基板各三件在含或不含额外牛骨髓脂肪条件下骨水泥固定。
- 所有模型行失效测试并测量轴向拔出分离力,记录植入物类型、牛骨髓有无和失效载荷并比较。
- 采用双因素方差分析及事后两两比较,统计学显著性设定为P < .05。
主要发现
- 第二代胫骨基板失效所需力显著更大。
- 牛骨髓存在整体显著降低各植入物设计的拔出分离力。
- 第一代与对照基板间未检测到显著差异。
- 各模型失效模式不同,与骨髓脂肪有无无关。
- 所有第一代基板均出现骨水泥-植入物界面脱粘。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Early clinical results of a new total knee arthroplasty (TKA) implant design show promise for improved outcomes and patellofemoral function scores. However, reports of early tibial component-cement interface debonding requiring revision have been published. This study investigated the biomechanical properties of three different tibial baseplates to understand potential causes of failure. METHODS: PFC Sigma (control), Attune (1st generation) and Attune S+ (2nd generation) tibial baseplates were implanted into 4th generation sawbone tibia models using a standardized technique. Three of each baseplate were cemented with and without additional bovine bone marrow fat. All models were tested to failure with measured axial distraction force. Implant type, presence or absence of bovine marrow and load to failure were all recorded and compared. Two-way ANOVA followed by post-hoc pairwise comparisons were used to determine statistical significance, which was set to P < .05. RESULTS: The 2nd generation tibial baseplates required significantly more force to failure. The presence of bovine marrow significantly reduced the pullout force of the implant designs overall. No significant difference was detected between the 1st generation and control baseplates. Failure mode for each model was also noted to be different irrespective of the presence or absence of bone marrow fat. CONCLUSION: The 2nd generation baseplates required significantly more force to failure compared with older designs. The presence of bone marrow during cementation of a tibial base plate significantly decreased axial pullout strength of a tibial baseplate in this laboratory model. All 1st generation baseplates exhibited debonding at the cement-implant interface.
引用
- Kelly BC, Owen JR, Shah SC, et al. A Biomechanical Comparison of the Effect of Baseplate Design and Bone Marrow Fat Infiltration on Tibial Baseplate Pullout Strength. J Arthroplasty. 2020. PMID: 32829970. doi:10.1016/j.arth.2020.07.049
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。