单次使用器械植入当代全膝关节置换系统准确、高效且安全
多中心研究比较单次使用器械(SUI)与可重复使用器械(RUI)在全膝关节置换术(TKA)中的植入准确性和手术室效率,共88例初级TKA(各44例),75例完成研究。SUI在力学轴对线绝对值上非劣于RUI,多数影像学参数无显著差异,但股骨假体屈曲角有轻微差异(P=0.015)。SUI手术室准备时间显著短于RUI(18.8分钟 vs 26.7分钟,P<0.001),手术时间无显著差异(P=0.295),无翻修且不良事件数无显著差异。
要点
- SUI在力学轴对线绝对值上非劣于RUI,多数影像学参数无显著差异。
- SUI手术室准备时间18.8分钟,短于RUI的26.7分钟(P<0.001)。
- 手术时间SUI 64.6分钟 vs RUI 60.5分钟,无显著差异(P=0.295)。
核心结论
- SUI植入物放置准确性与RUI相似,手术室准备时间缩短且不良事件未增加。
- 这些结果支持SUI用于TKA的安全性和有效性。
- 需要对潜在的经济和技术优势进行进一步分析。
研究设计与样本
- 多中心研究,四位外科医生共植入88例初级TKA,RUI组44例,SUI组44例。
- 术后3个月影像学测量准确性,主要终点为力学轴对线绝对值的非劣效性。
- 评估影像学终点、手术室时间和不良事件。
主要发现
- 75例完成研究(SUI组41例,RUI组34例)。
- SUI与RUI在多数影像学参数(股骨远端内翻-外翻、胫骨近端内翻-外翻、胫骨后倾或目标3°以内受试者比例)无显著差异。
- 股骨假体屈曲角有轻微差异(P=0.015)。
- SUI和RUI平均(标准差)手术室准备时间分别为18.8(10.03)分钟和26.7(6.93)分钟(P<0.001)。
- 手术时间(从第一切口到最后缝合)分别为64.6(16.95)分钟和60.5(19.01)分钟(P=0.295)。
局限性
- 需要对潜在的经济和技术优势进行进一步分析。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Single use instruments (SUI) is a potential mechanism to improve efficiency and reduce cost in total knee arthroplasty (TKA). New technology requires patient safety and surgical accuracy. A multi-center study of SUI vs reusable mechanical instrumentation (RUI) for a TKA system compared implant placement accuracy and operating room (OR) efficiency. METHODS: Four surgeons implanted 88 primary TKAs, N = 44 RUI and N = 44 SUI. Accuracy was measured radiographically at 3 months. The primary endpoint was non-inferiority of absolute value of mechanical axis alignment. Radiographic endpoints, OR times, and adverse events were also evaluated. RESULTS: Seventy-five subjects completed the study (41 SUI/34 RUI). The primary endpoint non-inferiority of SUI vs RUI was met, with no significant difference between SUI and RUI in most radiographic parameters (distal femoral varus-valgus, proximal tibial varus-valgus, tibial slope, or subjects within 3° of target); there was a slight difference in femoral component flexion angle (P = .015). SUI and RUI mean (SD) OR set-up times were 18.8 (10.03) and 26.7 (6.93) (P <.001), and surgical times (first incision to last stitch) were 64.6 (16.95) and 60.5 (19.01) (P = .295), respectively. Differences in OR clean-down and anesthesia were not significant. There were no revisions, and there was no significant difference in the number of reported adverse events. CONCLUSION: SUI resulted in similar accuracy of implant placement to RUI with decreased OR set-up time and no increase in adverse events. These results support the safety and efficacy of SUI for performing TKA. Further analysis of potential economic and technical advantages is warranted.