现代翻修股骨柄设计在下沉率方面无差异

Modern Revision Femoral Stem Designs Have No Difference in Rates of Subsidence.
PubMed 文献 · J Arthroplasty · 2020-08-05

研究回顾了335例翻修全髋关节置换术患者,比较模块化锥形开槽钛柄、一体式锥形开槽钛柄与全多孔涂层圆柱柄的影像学下沉率、再翻修率和患者报告结局。平均随访39个月,三组在平均下沉率、HOOS Jr评分和无菌性再翻修率上均无显著差异,多变量分析显示三组间下沉>5mm的风险也无差异。作者认为三种股骨柄设计在翻修全髋关节置换术中表现均良好,术者应选择其认为临床最合适的股骨柄。

要点

  • 平均随访39个月,模块化锥形开槽钛柄、一体式锥形开槽钛柄与全多孔涂层圆柱柄的平均下沉率分别为3.5mm、2.4mm、2.1mm,P=0.14。
  • 三组HOOS Jr评分分别为78分、74分、64分(P=0.15),无菌性再翻修率分别为4%、3%、0%(P=0.29)。
  • 多变量分析显示三种股骨柄间下沉>5mm的风险无差异(P>0.05)。

核心结论

  • 模块化锥形开槽钛柄、一体式锥形开槽钛柄和全多孔涂层圆柱柄在翻修全髋关节置换术中均表现良好,临床结局或下沉率无差异。
  • 作者认为术者应选择其认为临床最合适的股骨柄。

研究设计与样本

  • 回顾性连续队列研究,纳入2012年至2019年间接受股骨翻修的335例患者,其中模块化锥形开槽钛柄225例、一体式锥形开槽钛柄63例、全多孔涂层圆柱柄47例。
  • 评估影像学下沉率、再翻修率和患者报告结局,并行多变量回归分析确定股骨柄类型对下沉>5mm风险的独立影响。

主要发现

  • 平均随访39个月(范围12至96个月),模块化锥形开槽钛柄、一体式锥形开槽钛柄与全多孔涂层圆柱柄的平均下沉率分别为3.5mm、2.4mm、2.1mm(P=0.14)。
  • 三组HOOS Jr评分分别为78分、74分、64分(P=0.15),无菌性再翻修率分别为4%、3%、0%(P=0.29)。
  • 模块化锥形开槽钛柄更常用于股骨骨缺损严重(Paprosky III和IV型)的患者,但多变量分析显示三种股骨柄间下沉>5mm的风险无差异(P>0.05)。

临床怎么用

  • 作者认为术者应选择其认为临床最合适的股骨柄。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Both modular and monoblock tapered fluted titanium (TFT) stems have gained popularity over fully porous-coated cylindrical (FPCC) femoral stem designs, but limited data exist comparing subsidence rates following revision total hip arthroplasty (THA). The purpose of this study is to determine differences in subsidence and clinical outcomes among 3 revision femoral stem designs. METHODS: We reviewed a consecutive series of 335 patients who underwent femoral component revision to a cementless modular TFT (n = 225), monoblock TFT (n = 63), or FPCC (n = 47) stem between 2012 and 2019. We evaluated radiographic subsidence rates, re-revision rates, and patient-reported outcomes between the 3 stems. A multivariate regression analysis was performed to determine the independent effect of stem type on the risk of subsidence >5 mm. RESULTS: At an average follow-up of 39 months (range, 12 to 96 months), there were no differences in mean subsidence rates (3.5 vs 2.4 vs 2.1 mm, P = .14), HOOS Jr scores (78 vs 74 vs 64 points, P = .15), or aseptic re-revision rates (4% vs 3% vs 0%, P = .29) between modular TFT, monoblock TFT, and FPCC stems. Although modular TFT stems were more often used in patients with extensive femoral bone loss (Paprosky III and IV), there were no differences in subsidence rates >5 mm among the 3 stems (P > .05) in multivariate analysis. CONCLUSION: Modular TFT, monoblock TFT, and FPCC femoral stem designs all perform well in revision THA with no difference in clinical outcomes or subsidence rates. Surgeons should select the stem which they feel is the most clinically appropriate.