非骨水泥胫骨托在功能负荷条件下的微动与迁移
在7例尸体膝关节上植入市售非骨水泥全膝关节置换假体,经1050 N垂直负荷下0°-100°屈曲500次预适应后,施加行走和下楼时的峰值力与力矩,用数字图像相关法测量骨-假体界面三维运动。胫骨托预适应期平均迁移101±25 μm,下楼时后部区域微动显著高于行走,界面相对运动主要呈矢状面翘板样运动。
要点
- 预适应期胫骨托平均迁移101±25 μm,以矢状面旋转为主(占总迁移92%)。
- 下楼时5区和6区总微动显著高于行走(62±9 vs 51±10 μm;68±17 vs 37±10 μm,P<.05)。
- 下楼时前部区域倾斜(31±17 vs -16±20 μm)和后部区域倾斜(-60±8 vs -40±7 μm)均显著大于行走(P<.05)。
核心结论
- 非骨水泥胫骨托骨-假体界面相对运动在周边各区域差异明显,在评估的负荷条件下胫骨托主要呈矢状面翘板样运动。
- 与行走相比,下楼产生显著更多的微动,尤其在后部区域。
研究设计与样本
- 在7例尸体膝关节上植入市售非骨水泥全膝关节置换假体,在定制多轴功能活动模拟器上以1050 N垂直负荷行0°-100°屈曲500次预适应。
- 随后施加行走和下楼时的峰值力与力矩,用数字图像相关法测量骨-假体界面三维运动。
主要发现
- 预适应期胫骨托平均迁移101±25 μm,以矢状面旋转为主(占总迁移92%),合并后移(28%)和极小的横断面旋转(14%)。
- 6个测量区之间迁移差异达2.7倍(61-167 μm)。
- 5区总微动下楼62±9 μm vs 行走51±10 μm(P<.05),6区下楼68±17 μm vs 行走37±10 μm(P<.05)。
- 下楼时前部区域倾斜31±17 μm vs 行走-16±20 μm(P<.05),后部区域倾斜-60±8 μm vs 行走-40±7 μm(P<.05)。
- 下楼时前部区域前后位移-25±3 μm vs 行走-13±2 μm(P<.05)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The outcome of cementless total knee arthroplasty (TKA) relies on successful bony ingrowth into the implant surfaces. Failures due to aseptic loosening are still reported, especially in younger and more active patients. The objective of this study is to quantify the micromotion of a commercially available design of cementless tibial tray under loading conditions simulating walking and stair descent. METHOD: A commercially available design of cementless total knee arthroplasty was implanted in 7 cadaveric knees which were preconditioned with 500 cycles of 0°-100° flexion under a vertical load of 1050 N in a custom-built, multiaxial functional activity simulator. This was followed by application of the peak forces and moments occurring during walking and stair descent. During each loading procedure, 3-dimensional motion at the bone-prosthesis interface was measured using digital image correlation. RESULTS: The tray migrated 101 ± 25 μm on average during preconditioning, which was dominated by rotation in the sagittal plane (92% of total migration), combined with posterior translation (28%) and minimal rotation in the transverse plane (14%). The migration varied 2.7-fold (61-167 μm) between the 6 measurement zones. Stair descent produced significantly higher total micromotion than walking in zone #5 (62 ± 9 vs 51 ± 10 μm, P < .05) and zone #6 (68 ± 17 vs 37 ± 10 μm, P < .05). In addition, during stair descent, the tray exhibited significantly more tilting (anterior zones: 31 ± 17 vs -16 ± 20 μm, P < .05; posterior zones: -60 ± 8 vs -40 ± 7 μm, P < .05) and more anteroposterior displacement in the anterior zones (-25 ± 3 vs -13 ± 2 μm, P < .05) when compared to walking. CONCLUSION: The relative motion at the bone-prosthesis interface varied substantially around the periphery of the cementless tray. Under the loading conditions evaluated, the tray primarily underwent a rocking motion in the sagittal plane. Compared with walking, stair descent produced significantly more micromotion, especially in the posterior zones.
引用
- Han S, Patel RV, Ismaily SK, et al. Micromotion and Migration of Cementless Tibial Trays Under Functional Loading Conditions. J Arthroplasty. 2020. PMID: 32741712. doi:10.1016/j.arth.2020.07.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。