一体式双动髋臼杯全髋关节置换术治疗老年股骨颈骨折手术固定并发症发生率高
这项前瞻性研究纳入97例(97髋)移位股骨颈骨折老年患者,平均年龄76.6岁,平均股骨颈骨密度T值为-2.8,均行一体式双动髋臼杯全髋关节置换术。术后CT评估显示臼杯外倾角异常占10.3%(10/97)、前倾角异常占35.1%(34/97),髋臼假体周围骨折6例(6.2%),再手术率4.1%(4/97)。作者认为,在骨量差的老年患者中,一体式双动髋臼杯与臼杯固定不当及髋臼假体周围骨折相关,尽管双动髋臼杯降低了全髋关节置换术后早期脱位风险。
要点
- 97例老年股骨颈骨折患者行一体式双动髋臼杯全髋关节置换术。
- 臼杯外倾角异常10.3%(10/97),前倾角异常35.1%(34/97)。
- 髋臼假体周围骨折6例(6.2%),再手术率4.1%(4/97)。
核心结论
- 在骨量差的老年股骨颈骨折患者中,使用一体式双动髋臼杯与臼杯固定不当及髋臼假体周围骨折相关,尽管双动髋臼杯降低了全髋关节置换术后早期脱位风险。
研究设计与样本
- 前瞻性研究,纳入97例(97髋)移位股骨颈骨折并知情同意参加的患者,平均年龄76.6岁(范围60-95岁),平均股骨颈骨密度T值为-2.8(范围-1.2至-5.5)。
- 所有患者均行一体式双动髋臼杯全髋关节置换术,并行计算机断层扫描(CT)评估髋臼杯前倾角、外倾角、松动及髋臼假体周围骨折等影像学参数。
主要发现
- 臼杯平均外倾角为40.2°(范围23.5°-63°),平均前倾角为32.6°(范围7°-66.2°)。
- 手术位置异常比例在外倾角为10.3%(10/97),在前倾角为35.1%(34/97)。
- 2周内早期臼杯松动检出2髋。
- 髋臼假体周围骨折检出6髋(6/97,6.2%),其中5例无移位骨折经保守治疗愈合,1例移位骨折最终导致臼杯松动并接受翻修手术。
- 一体式双动髋臼杯再手术率为4.1%(4/97)。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: This study is prospectively conducted to evaluate surgical complications of monolithic dual mobility cup total hip arthroplasty (THA) in elderly patients with fractured neck of the femur. METHODS: Ninety-seven patients (97 hips) with displaced femoral neck fracture who gave informed consent for participation were prospectively enrolled. Their mean age was 76.6 years (range, 60-95 years), and the mean bone mineral density T-score of neck of the femur was -2.8 (range, -1.2 to -5.5). All patients underwent THA with monolithic dual mobility cup, and computed tomography scans were obtained to evaluate radiographic parameters including anteversion, inclination, and loosening of acetabular cups, and periprosthetic acetabular fractures. RESULTS: With regard to cup orientation, mean inclination angle was 40.2° (range, 23.5°-63°) and mean anteversion was 32.6° (range, 7°-66.2°). The proportion of surgical outliers was 10.3% (10/97) in inclination and 35.1% (34/97) in anteversion. Early cup loosening within 2 weeks was detected in 2 hips. Periprosthetic acetabular fractures were identified in 6 hips (6/97, 6.2%). Of the 6 fractures, 5 nondisplaced fractures were healed with conservative management, but 1 fracture with displacement eventually led to cup loosening and the patient underwent revision surgery. Reoperation rate of the monolithic dual mobility cup was 4.1% (4/97). CONCLUSION: The use of the monolithic dual mobility was associated with improper cup fixation and periprosthetic acetabular fractures in the elderly with poor bone stock, although the dual mobility cup lowered the risk of early dislocation after THA.
引用
- Sunilkumar PD, Oh KJ, Cho HW, et al. Monolithic Dual Mobility Cup Total Hip Arthroplasty Has High Complication Rates With Surgical Fixation in Elderly With Femur Neck Fracture. J Arthroplasty. 2020. PMID: 32773273. doi:10.1016/j.arth.2020.06.091
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。