文献速递关节外科创伤骨科

股骨距领可预防早期扭转/螺旋形假体周围股骨骨折:配对尸体生物力学分析

A Calcar Collar Is Protective Against Early Torsional/Spiral Periprosthetic Femoral Fracture: A Paired Cadaveric Biomechanical Analysis.
PubMed 文献 · J Bone Joint Surg Am · 2020-01-01

该研究采用11对配对尸体股骨(平均年龄77±10.2岁)进行生物力学测试,比较带领与无领股骨柄在模拟68 kg体重负荷下抵抗扭转至骨折的峰值扭矩。结果显示带领股骨柄的中位骨折扭矩为65.4 Nm,显著高于无领股骨柄的43.1 Nm(p=0.0014,Wilcoxon符号秩检验),骨折模式均为假体周围螺旋形骨折。作者认为带领股骨柄在扭转负荷下似乎具有保护作用。

要点

  • 带领股骨柄中位骨折扭矩65.4 Nm,无领股骨柄43.1 Nm,差异有统计学意义(p=0.0014)。
  • 所有标本的失败模式均为假体周围螺旋形骨折。
  • 带领股骨柄在扭转负荷下似乎具有保护作用,可能转化为预防早期假体周围Vancouver B2型股骨骨折。

核心结论

  • 在这一配对股骨生物力学模型中,带领股骨柄在扭转负荷下似乎具有保护作用。
  • 作者认为这些结果可能转化为预防骨整合发生前的早期假体周围Vancouver B2型股骨骨折。

研究设计与样本

  • 研究采用22根尸体股骨(11对配对),平均年龄77±10.2岁(范围54至90岁)。
  • 每对中一根股骨植入带领股骨柄,对侧植入无领股骨柄,两者具有相同的骨内几何形状。
  • 采用标准扩髓技术植入,施加68 kg压缩负荷模拟行走时体重,再施加旋转位移直至骨折。
  • 使用扭矩计载荷传感器和数据采集软件测量骨折前峰值扭矩。
  • 采用Wilcoxon符号秩检验进行统计分析。

主要发现

  • 带领股骨柄的中位骨折扭矩为65.4 Nm,无领股骨柄为43.1 Nm(p=0.0014,Wilcoxon符号秩检验)。
  • 中位T值为-1.95(范围-4.1至-0.15)。
  • 骨折扭矩的中位差值为29.18 Nm。
  • 每例的失败模式均为围绕植入物的螺旋形骨折。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Periprosthetic fracture is a leading reason for readmission following total hip arthroplasty. Most of these fractures occur during the early postoperative period before bone ingrowth. Before ingrowth occurs, the femoral component can rotate relative to the femoral canal, causing a spiral fracture pattern. We sought to evaluate, in a paired cadaver model, whether the torsional load to fracture was higher in collared stems. The hypothesis was that collared stems have greater load to fracture under axial and torsional loads compared with collarless stems. METHODS: Twenty-two cadaveric femora (11 matched pairs) with a mean age of 77 ± 10.2 years (range, 54 to 90 years) were harvested. Following dissection, the femora were evaluated with use of a dual x-ray absorptiometry scanner and T scores were recorded. We utilized a common stem that is available with the same intraosseous geometry with and without a collar. For each pair, 1 femur was implanted with a collared stem and the contralateral femur was implanted with a collarless stem with use of a standard broaching technique. A compressive 68-kg load was applied to simulate body weight during ambulation. A rotational displacement was then applied until fracture occurred. Peak torque prior to fracture was measured with use of a torque meter load cell and data acquisition software. RESULTS: The median torque to fracture was 65.4 Nm for collared stems and 43.1 Nm for uncollared stems (p = 0.0014, Wilcoxon signed-rank test). The median T score was -1.95 (range, -4.1 to -0.15). The median difference in torque to fracture was 29.18 Nm. As expected in each case, the mode of failure was a spiral fracture around the implant. CONCLUSIONS: Collared stems seemed to offer a protective effect in torsional loading in this biomechanical model comparing matched femora. CLINICAL RELEVANCE: These results may translate into a protective effect against early periprosthetic Vancouver B2 femoral fractures that occur before osseous integration has occurred.