直接前入路全髋关节置换术中两种术中透视网格技术对假体放置准确性的比较

Comparison of Component Placement Accuracy Using Two Intraoperative Fluoroscopic Grid Technologies During Direct Anterior Total Hip Arthroplasty.
PubMed 文献 · J Arthroplasty · 2020-06-24

该回顾性研究比较了直接前入路全髋关节置换术中手动网格系统(MGS)与数字网格系统(DGS)辅助透视对髋臼杯外展角、双下肢不等长及整体髋偏移差的影响,发现两组均能获得较高的目标区放置率,MGS组外展角更接近目标且整体髋偏移差更小,但作者认为这些差异不太可能具有临床意义。

要点

  • MGS组与DGS组分别有98.7%和96.8%的病例髋臼杯外展角达到目标范围(45°±10°)。
  • MGS组平均外展角为45.14°±4.03°,显著低于DGS组的47.01°±4.39°(P<.001)。
  • DGS组平均整体髋偏移差显著高于MGS组(4.45±2.73 mm vs 3.64±2.44 mm,P=.002),双下肢不等长无显著差异。

核心结论

  • 使用MGS或DGS均能使假体放置保持在预设目标区内,MGS似乎略更一致,但这些差异不太可能具有临床意义。

研究设计与样本

  • 回顾性评估连续患者的术后6周负重X线片,比较直接前入路全髋关节置换术中辅助使用MGS或DGS的髋臼杯外展角、双下肢不等长和整体髋偏移差,采用独立样本t检验分析组间差异。
  • MGS组250例患者(315髋),DGS组183例患者(218髋)。

主要发现

  • MGS组和DGS组分别有98.7%和96.8%的病例髋臼杯外展角达到目标范围(45°±10°),MGS组平均外展角显著更低(45.14°±4.03° vs 47.01°±4.39°,P<.001)。
  • DGS组平均整体髋偏移差显著高于MGS组(4.45±2.73 mm vs 3.64±2.44 mm,P=.002)。
  • 两组双下肢不等长无显著差异(MGS组2.92±2.55 mm vs DGS组3.19±2.46 mm,P=.275)。
  • 达到目标双下肢不等长和整体髋偏移差百分比两组间无显著差异,但同时达到全部三项目标的比例MGS组为97.6%,DGS组为93.5%(P=.031)。

关键数值与推荐

  • 预设目标:双下肢不等长<10 mm,整体髋偏移差<10 mm,髋臼杯外展角45°±10°。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Intraoperative fluoroscopy is beneficial when performing total hip arthroplasty (THA) via the direct anterior approach; however, image distortion may influence component placement. A manual gridding system (MGS) and a digital gridding system (DGS) are commercially available, aimed at visually representing or correcting image distortion. Therefore, the purpose of this study is to compare component placement accuracy following direct anterior approach THA when intraoperative fluoroscopy was supplemented with MGS or DGS. METHODS: A retrospective evaluation of acetabular cup abduction (ABD), leg length discrepancy (LLD) and global hip offset difference (GHO) was completed for consecutive patients from 6 week post-THA weight-bearing radiographs. The predefined target LLD and GHO was <10 mm and ABD target was 45° ± 10°. Differences between MGS and DGS were determined by independent t-tests. RESULTS: The MGS (250 patients, 315 hips) and DGS (183 patients, 218 hips) achieved targeted ABD in 98.7% and 96.8% of cases, respectively, and ABD was significantly lower in the MGS group (45.14 ± 4.03° and 47.01 ± 4.39°, respectively) (P < .001). Compared to MGS, the DGS group averaged significantly higher GHO (3.64 ± 2.44 and 4.45 ± 2.73 mm, respectively, P = .002) but was not significantly different regarding LLD (2.92 ± 2.55 and 3.19 ± 2.46 mm, respectively, P = .275). No significant group difference was noted for percentage within the targeted LLD and GHO; however, 93.5% of DGS and 97.6% of MGS achieved all three (P = .031). CONCLUSION: The use of both the MGS and DGS resulted in consistent component placement within the predefined target zone. Although the MGS appeared to be slightly more consistent, these differences are unlikely to be clinically significant.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Thorne TJ, Nishioka ST, Andrews SN, et al. Comparison of Component Placement Accuracy Using Two Intraoperative Fluoroscopic Grid Technologies During Direct Anterior Total Hip Arthroplasty. J Arthroplasty. 2020. PMID: 32680756. doi:10.1016/j.arth.2020.06.053