2020年Otto Aufranc奖:模块化双动内衬安装不良

2020 Otto Aufranc Award: Malseating of modular dual mobility liners.
PubMed 文献 · Bone Joint J · 2020-01-01

该研究结合临床与体外实验,在551例初次全髋关节置换术(THA)中评估模块化双动(MDM)内衬安装不良的发生率,并利用模拟腐蚀腔测试安装不良对微动腐蚀的影响。结果显示32例(5.8%)发生安装不良,低手术量医生发生率显著更高(p < 0.001);安装不良组在2400 N下微动起始载荷更低,提示可能易发生微动腐蚀。

要点

  • 551例MDM初次THA中32例(5.8%)影像学诊断内衬安装不良,三种杯设计均有发生。
  • 低MDM手术量医生安装不良发生率显著高于高手术量医生(p < 0.001)。
  • 体外模拟腐蚀腔中,安装不良组在2400 N下微动起始载荷更低;良好安装组在大于800 N的所有峰值载荷下微动电流显著更低(p < 0.044)。

核心结论

  • MDM内衬安装不良在高容量关节置换中心至少每20例患者中可发生1例。
  • 加载周期中微动起始和微动电流增加提示安装不良时可能易发生腐蚀,支持安装不良内衬可能存在微动腐蚀风险的假说。
  • 临床医生应了解这一现象。

研究设计与样本

  • 临床部分:观察者独立审阅同一制造商MDM初次THA术后X线片551例,时间跨度三年,采用多变量logistic回归分析年龄、性别、BMI、杯设计、杯尺寸和医生MDM手术量等危险因素。
  • 体外部分:6个含钴铬内衬的未使用MDM植入物在模拟腐蚀腔中测试,3个良好安装,3个以6°倾斜安装不良,进行递增幅度循环加载并测量微动电流。

主要发现

  • 551例MDM内衬中32例(5.8%)影像学诊断安装不良,三种杯设计均有发生。
  • 低MDM手术量医生安装不良发生率显著高于高MDM手术量医生(p < 0.001)。
  • 体外测试中,良好安装的未使用内衬在大于800 N的所有峰值载荷下微动电流值显著更低(p < 0.044)。
  • 安装不良的内衬-外壳配对在2400 N下微动起始载荷更低。

临床怎么用

  • 临床医生应了解MDM内衬安装不良这一现象。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: This combined clinical and in vitro study aimed to determine the incidence of liner malseating in modular dual mobility (MDM) constructs in primary total hip arthroplasties (THAs) from a large volume arthroplasty centre, and determine whether malseating increases the potential for fretting and corrosion at the modular metal interface in malseated MDM constructs using a simulated corrosion chamber. METHODS: For the clinical arm of the study, observers independently reviewed postoperative radiographs of 551 primary THAs using MDM constructs from a single manufacturer over a three-year period, to identify the incidence of MDM liner-shell malseating. Multivariable logistic regression analysis was performed to identify risk factors including age, sex, body mass index (BMI), cup design, cup size, and the MDM case volume of the surgeon. For the in vitro arm, six pristine MDM implants with cobalt-chrome liners were tested in a simulated corrosion chamber. Three were well-seated and three were malseated with 6° of canting. The liner-shell couples underwent cyclic loading of increasing magnitudes. Fretting current was measured throughout testing and the onset of fretting load was determined by analyzing the increase in average current. RESULTS: The radiological review identified that 32 of 551 MDM liners (5.8%) were malseated. Malseating was noted in all of the three different cup designs. The incidence of malseating was significantly higher in low-volume MDM surgeons than high-volume MDM surgeons (p < 0.001). Pristine well-seated liners showed significantly lower fretting current values at all peak loads greater than 800 N (p < 0.044). Malseated liner-shell couples had lower fretting onset loads at 2,400 N. CONCLUSION: MDM malseating remains an issue that can occur in at least one in 20 patients at a high-volume arthroplasty centre. The onset of fretting and increased fretting current throughout loading cycles suggests susceptibility to corrosion when this occurs. These results support the hypothesis that malseated liners may be at risk for fretting corrosion. Clinicians should be aware of this phenomenon. Cite this article: Bone Joint J 2020;102-B(7 Supple B):20-26.