全踝关节置换胫骨垫片冠状面力线不良对远端足骨生物力学的影响:尸体步态研究
该尸体步态模拟研究在7例足标本中植入改良全踝关节置换(TAA),用楔形垫模拟胫骨垫片5°、10°、15°内翻和外翻,量化踝关节一致性与足部运动学及足底压力变化。结果显示内翻力线使胫距关节内翻并内旋、活动度轻度增加而第一跖趾关节活动度轻度减小;外翻力线主要增加拇趾压力并轻度减少第三、四跖骨压力。
要点
- TAA胫骨垫片冠状面内翻力线可致胫距关节内翻、内旋及活动度轻度增加。
- 外翻力线主要表现为拇趾压力增加及第三、四跖骨轻度减载。
- 该尸体模型有助于理解力线不良与临床不良结局相关的生物力学机制。
核心结论
- 在尸体步态模拟中,TAA胫骨垫片冠状面力线不良会改变足部运动学与足底压力。
- 内翻力线主要导致胫距关节内翻、内旋、胫距关节活动度轻度增加及第一跖趾关节活动度轻度减小。
- 外翻力线主要表现为拇趾压力增加,并伴第三、四跖骨压力轻度降低。
研究设计与样本
- 研究在7例尸体足标本中植入改良TAA,使用楔形垫模拟胫骨垫片冠状面力线不良。
- 通过影像学量化中立位及5°、10°、15°内翻和外翻楔形垫下的胫骨垫片角(TIA)与距骨组件角(TCA)。
- 以机器人步态模拟器模拟1/6生理速度行走,运动捕捉系统测足部运动学,压力垫测足底压力。
- 关节一致性以TIA与TCA差值量化,采用线性混合效应回归比较各力线不良条件与中立位。
主要发现
- 描述性显示TIA与TCA均小于楔形角度,且存在组件不一致(R=0.65;p<0.001)。
- 胫骨垫片内翻力线使胫距关节内翻并内旋,胫距关节活动度轻度增加(1.3±0.7°[95%CI -0.7至3.4];p=0.03),第一跖趾关节活动度减小(-1.9±0.9°[95%CI -5.6至1.7];p=0.007)。
- 矢状面舟楔关节活动度随TIA内翻轻度减小(-0.9±0.4°[95%CI -2.1至0.3];p=0.017)。
- 拇趾压力随TIA外翻增加(59±50 kPa[95%CI -88至207];p=0.006)。
- 第三、四跖骨峰值压力随TIA外翻轻度降低(-15±17°[95%CI -65至37];p=0.03;-8±4°[95%CI -17至1];p=0.048),第五跖骨压力在外翻(-18±12 kPa[95%CI -51至15])或内翻(-7±18 kPa[95%CI -58至45])时轻度降低(p=0.002);所有比较均与中立位对照。
临床怎么用
- TAA术中应关注胫骨垫片冠状面力线,内翻与外翻力线不良可产生不同的足部生物力学改变。
- 内翻力线不良可能影响胫距关节及第一跖趾关节运动学,外翻力线不良可能增加拇趾负荷。
- 该结果可为理解力线不良相关临床不良结局(如患者报告结局差或假体松动)提供生物力学参考。
局限性
- 研究为尸体步态模拟,行走速度为1/6生理速度,可能不完全反映生理步态。
- 样本量为7例尸体足标本,结果外推需谨慎。
- 使用楔形垫模拟力线不良,可能无法完全代表临床TAA力线不良的复杂情况。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Total ankle arthroplasty (TAA) is becoming a more prevalent treatment for end-stage ankle arthritis. However, the effects of malalignment on TAA remain poorly understood. QUESTIONS/PURPOSES: The purpose of this study was to quantify the mechanical effects of coronal plane malalignment of the tibial insert in TAA using cadaveric gait simulation. Specifically, we asked, is there a change in (1) ankle joint congruency, (2) kinematic joint position, (3) kinematic ROM, (4) peak plantar pressure, and (5) center of pressure with varus and valgus malalignment? METHODS: A modified TAA was implanted into seven cadaveric foot specimens. Wedges were used to simulate coronal plane malalignment of the tibial insert. The degree of malalignment (tibial insert angle [TIA] and talar component angle [TCA]) was quantified radiographically for neutral and 5°, 10°, and 15° varus and valgus wedges. Dynamic walking at 1/6 of physiological speed was simulated using a robotic gait simulator. A motion capture system was used to measure foot kinematics, and a pressure mat was used to measure plantar pressure. Joint congruency was quantified as the difference between TIA and TCA. Continuous joint position, joint ROM, peak plantar pressure, and center of pressure for varus and valgus malalignment compared with neutral alignment were estimated using linear mixed effects regression. Pairwise comparisons between malalignment conditions and neutral were considered significant if both the omnibus test for the overall association between outcome and malalignment and the individual pairwise comparison (adjusted for multiple comparisons within a given outcome) had p ≤ 0.05. RESULTS: Descriptively, the TIA and TCA were both less pronounced than the wedge angle and component incongruence was seen (R = 0.65; p < 0.001). Varus malalignment of the tibial insert shifted the tibiotalar joint into varus and internally rotated the joint. The tibiotalar joint's ROM slightly increased as the TIA shifted into varus (1.3 ± 0.7° [mean ± SD] [95% confidence interval -0.7 to 3.4]; p = 0.03), and the first metatarsophalangeal joint's ROM decreased as the TIA shifted into varus (-1.9 ± 0.9° [95% CI -5.6 to 1.7]; p = 0.007). In the sagittal plane, the naviculocuneiform joint's ROM slightly decreased as the TIA shifted into varus (-0.9 ± 0.4° [95% CI -2.1 to 0.3]; p = 0.017). Hallux pressure increased as the TIA became more valgus (59 ± 50 kPa [95% CI -88 to 207]; p = 0.006). The peak plantar pressure slightly decreased in the third and fourth metatarsals as the TIA shifted into valgus (-15 ± 17° [95% CI -65 to 37]; p = 0.03 and -8 ± 4° [95% CI -17 to 1]; p = 0.048, respectively). The fifth metatarsal's pressure slightly decreased as the TIA shifted into valgus (-18 ± 12 kPa [95% CI -51 to 15]) or varus (-7 ± 18 kPa [95% CI -58 to 45]; p = 0.002). All comparisons were made to the neutral condition. CONCLUSIONS: In this cadaver study, coronal plane malalignment in TAA altered foot kinematics and plantar pressure. In general, varus TAA malalignment led to varus shift and internal rotation of the tibiotalar joint, a slight increase in the tibiotalar ROM, and a slight decrease in the first metatarsophalangeal ROM, while a valgus TAA malalignment was manifested primarily through increased hallux pressure with a slight off-loading of the third and fourth metatarsals. CLINICAL RELEVANCE: This study may increase our understanding of the biomechanical processes that underlie the unfavorable clinical outcomes (such as, poor patient-reported outcomes or implant loosening) that have been associated with coronal plane malalignment of the tibial component in TAA.
引用
- Buckner BC, Stender CJ, Baron MD, et al. Does Coronal Plane Malalignment of the Tibial Insert in Total Ankle Arthroplasty Alter Distal Foot Bone Mechanics? A Cadaveric Gait Study. Clin Orthop Relat Res. 2020. PMID: 32574472. doi:10.1097/CORR.0000000000001294
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。