文献速递关节外科

后稳定型全膝关节置换中非骨水泥与骨水泥胫骨固定的随机试验

Cementless Versus Cemented Tibial Fixation in Posterior Stabilized Total Knee Replacement: A Randomized Trial.
PubMed 文献 · J Bone Joint Surg Am · 2020-01-01

这项单中心盲法随机试验纳入100例后稳定型全膝关节置换患者,比较骨水泥与非骨水泥胫骨固定的组件迁移、骨重塑和临床结局。24个月时非骨水泥组下沉和最大总点运动显著大于骨水泥组,骨水泥组假体周围骨密度丢失更多,临床评分两组改善相同。作者建议ACS后稳定型全膝关节置换使用骨水泥胫骨固定。

要点

  • 非骨水泥胫骨固定出现晚期持续迁移,24个月下沉1.22 mm,骨水泥组仅0.06 mm。
  • 24个月最大总点运动非骨水泥组2.04 mm,骨水泥组0.48 mm,差异显著。
  • 骨水泥组在内侧、外侧和前侧假体周围区域骨密度丢失多于非骨水泥组。

核心结论

  • 非骨水泥胫骨固定出现晚期持续下沉和高最大总点运动,提示ACS后稳定型膝假体固定稳定性存疑。
  • 骨水泥胫骨组件稳定,作者建议ACS后稳定型全膝关节置换采用骨水泥胫骨固定。

研究设计与样本

  • 前瞻性单中心盲法随机临床试验,纳入100例平均年龄68岁(范围45至87岁)的患者。
  • 由2名经验丰富的关节置换外科医生使用ACS假体(Implantcast)手术,随机分为骨水泥或非骨水泥胫骨固定,其他变量标准化。

主要发现

  • 24个月平均下沉非骨水泥组1.22 mm,骨水泥组0.06 mm(p < 0.01)。
  • 24个月平均最大总点运动非骨水泥组2.04 mm,骨水泥组0.48 mm(p < 0.01)。
  • 骨水泥组在内侧、外侧和前侧假体周围区域骨密度丢失多于非骨水泥组。
  • 非骨水泥组2例胫骨组件翻修,其中1例因松动。
  • 两组临床评分(牛津膝关节评分和SF-12)改善相同。

关键数值与推荐

  • 采用放射立体测量分析(RSA)和双能X线吸收测定(DXA)在住院期间及3、12、24个月监测胫骨组件迁移和假体周围骨密度。

临床怎么用

  • 针对ACS后稳定型全膝关节置换,建议使用骨水泥胫骨固定以降低迁移风险。
  • 对非骨水泥胫骨固定患者需警惕晚期下沉和松动,加强影像学随访。

局限性

  • 单中心研究,仅使用ACS假体,结论外推需谨慎。
  • 样本量100例,随访24个月,长期结果未报告。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Evidence supports the use of cementless tibial component fixation in total knee replacement but there is a paucity of literature on posterior stabilized designs. This randomized study of posterior stabilized total knee replacements compared cemented and cementless tibial fixation with regard to component migration, bone remodeling, and clinical outcomes. METHODS: This prospective single-center, blinded, randomized clinical trial included 100 patients with a mean age of 68 years (range, 45 to 87 years). Operations were performed by either of 2 experienced arthroplasty surgeons using the Advanced Coated System (ACS) prosthesis (Implantcast). Patients were randomized to cemented or cementless tibial fixation; other variables were standardized. Radiostereometric analysis (RSA) and dual x-ray absorptiometry (DXA) were performed during the inpatient stay and at 3, 12, and 24 months to monitor tibial component migration and periprosthetic bone mineral density (BMD). Clinical scores including the Oxford Knee Score and Short Form-12 were assessed prior to surgery and during follow-up. RESULTS: Late and continuing migration was found with cementless fixation. At 24 months, the mean subsidence was significantly greater for the cementless group (1.22 mm) compared with the cemented group (0.06 mm) (p < 0.01). The mean maximum total point motion at 24 months was 2.04 mm for the cementless group and 0.48 mm for the cemented group (p < 0.01). The cemented group had more BMD loss than the cementless group in the medial, lateral, and anterior periprosthetic regions. Two tibial components were revised in the cementless group, including 1 for loosening. Clinical scores improved equally in the 2 groups. CONCLUSIONS: Late ongoing subsidence and high maximum total point motion in our patients who underwent cementless tibial fixation raise concerns about the fixation stability of ACS cementless posterior stabilized knee replacements. Cemented tibial components were stable. Thus, we advise caution regarding the use of cementless tibial components and recommend tibial fixation with cement for the ACS posterior stabilized total knee replacement. LEVEL OF EVIDENCE: Therapeutic Level II. See Instructions for Authors for a complete description of levels of evidence.

来源:J Bone Joint Surg Am,查看原文。版权归原作者所有。

引用

  1. Nivbrant NO, Khan RJK, Fick DP, et al. Cementless Versus Cemented Tibial Fixation in Posterior Stabilized Total Knee Replacement: A Randomized Trial. J Bone Joint Surg Am. 2020. PMID: 32559052. doi:10.2106/JBJS.19.01010

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。