庆大霉素缓释超高分子量聚乙烯间隔器治疗假体周围感染
研究提出一种由钴铬股骨组件和庆大霉素缓释超高分子量聚乙烯(UHMWPE)胫骨垫片组成的新型临时间隔器,用于替代传统骨水泥间隔器。体外实验显示该间隔器可根除浮游金黄色葡萄球菌并防止股骨和胫骨组件定植,且机械和磨损性能远优于传统骨水泥间隔器。结论认为其抗菌效果与骨水泥间隔器相当,同时克服了后者缺点,有望减少并发症并改善患者生活质量。
要点
- 庆大霉素缓释UHMWPE间隔器可根除浮游金黄色葡萄球菌并防止组件定植。
- 该间隔器机械和磨损性能远优于传统庆大霉素骨水泥间隔器。
- 抗菌效果与骨水泥间隔器相当,有望减少并发症并改善生活质量。
核心结论
- 庆大霉素缓释UHMWPE间隔器抗菌效果与现行骨水泥间隔器相当,同时克服了骨水泥间隔器的缺点。
- 该新型间隔器有潜力大幅减少与骨水泥间隔器相关的并发症,并显著改善治疗期间患者的生活质量。
研究设计与样本
- 采用相分离压缩模塑法制备庆大霉素负载UHMWPE,并与传统庆大霉素负载骨水泥对比药物洗脱动力学、抗菌、机械和磨损性能。
主要发现
- 庆大霉素负载UHMWPE胫骨组件不仅根除浮游金黄色葡萄球菌,还防止了股骨和胫骨组件的定植。
- 与常规骨水泥间隔器相比,所提出的间隔器具有远更优越的机械和磨损性能。
临床怎么用
- 该间隔器设计为替换传统骨水泥间隔器,用于假体周围感染的治疗,提供庆大霉素治疗性释放并保留优良机械性能。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: We propose a state-of-the-art temporary spacer, consisting of a cobalt-chrome (CoCr) femoral component and a gentamicin-eluting ultra-high molecular weight polyethylene (UHMWPE) tibial insert, which can provide therapeutic delivery of gentamicin, while retaining excellent mechanical properties. The proposed implant is designed to replace conventional spacers made from bone cement. METHODS: Gentamicin-loaded UHMWPE was prepared using phase-separated compression moulding, and its drug elution kinetics, antibacterial, mechanical, and wear properties were compared with those of conventional gentamicin-loaded bone cement. RESULTS: Gentamicin-loaded UHMWPE tibial components not only eradicated planktonic Staphylococcus aureus, but also prevented colonization of both femoral and tibial components. The proposed spacer possesses far superior mechanical and wear properties when compared with conventional bone cement spacers. CONCLUSION: The proposed gentamicin-eluting UHMWPE spacer can provide antibacterial efficacy comparable with currently used bone cement spacers, while overcoming their drawbacks. The novel spacer proposed here has the potential to drastically reduce complications associated with currently used bone cement spacers and substantially improve patients' quality of life during the treatment. Cite this article: Bone Joint J 2020;102-B(6 Supple A):151-157.
引用
- Gil D, Atici AE, Connolly RL, et al. Addressing prosthetic joint infections via gentamicin-eluting UHMWPE spacer. Bone Joint J. 2020. PMID: 32475290. doi:10.1302/0301-620X.102B6.BJJ-2019-1593.R1