文献速递脊柱外科小儿骨科

智能自动调压矫形器治疗青少年特发性脊柱侧凸:双中心随机对照试验

The Intelligent Automated Pressure-Adjustable Orthosis for Patients With Adolescent Idiopathic Scoliosis: A Bi-Center Randomized Controlled Trial.
PubMed 文献 · Spine (Phila Pa 1976) · 2020-01-01

结果显示PO组目标压力内佩戴质量提高33.9%,1年内无进展,而CO组有2例进展超过5°,两组生活质量无显著差异。

要点

  • PO组目标压力内佩戴质量比CO组高33.9%(56.5±16.5% vs 23.1±12.1%,P<0.001)。
  • 1年随访PO组无患者进展,CO组有2例进展超过5°。
  • PO组日均佩戴时间比CO组长1.1小时(15.4±4.5 vs 14.3±3.8 h)。

核心结论

  • 自动调压矫形器可提高佩戴质量,在观察期内提供更好的生物力学矫正效果,且不负面影响佩戴时长和生活质量。

研究设计与样本

  • 双中心随机对照试验,纳入符合年龄10-14岁、Cobb角20°-40°、Risser征0-2标准的青少年特发性脊柱侧凸患者,随机分为PO组和CO组。
  • 在基线、矫形器配戴后即刻和1年随访时进行影像学和生活质量评估,并通过温度和压力传感器追踪佩戴依从性。

主要发现

  • PO组即刻矫形内矫正11.0°±6.5°(42.0%,P<0.001),CO组10.3°±5.3°(37.6%,P<0.001)。
  • 1年后PO组无患者进展,CO组有2例进展超过5°。
  • PO组日均佩戴时间15.4±4.5小时,CO组14.3±3.8小时,PO组长1.1小时。
  • PO组目标压力内佩戴质量56.5±16.5%,CO组23.1±12.1%,PO组高33.9%(P<0.001)。
  • 两组间及组内生活质量结果在研究期间均无显著差异。

关键数值与推荐

  • 纳入标准:年龄10-14岁、Cobb角20°-40°、Risser征0-2。
  • 进展定义为超过5°。
  • 目标压力内佩戴质量PO组56.5±16.5% vs CO组23.1±12.1%(P<0.001)。

临床怎么用

  • 对于符合纳入标准的青少年特发性脊柱侧凸患者,自动调压矫形器可提高目标压力内佩戴质量并可能减少进展。
  • 自动调压矫形器未影响患者佩戴时长和生活质量,可作为提高矫形器佩戴质量的选项。

局限性

  • 样本量较小,最终分析仅23例(PO组11例、CO组12例)。
  • 随访时间为1年,长期效果未报告。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Randomized controlled trial. OBJECTIVE: To compare the effectiveness of the automated pressure-adjustable orthosis (PO) and conventional orthosis (CO) for treatment of adolescent idiopathic scoliosis (AIS). SUMMARY OF BACKGROUND DATA: Orthosis wearing quality may influence its effectiveness for AIS. An automated PO aimed to provide a more optimized and consistent biomechanical environment. Clinical evaluation was conducted to study the effectiveness of this innovative orthosis. METHODS: Patients with AIS who met the criteria (Age 10-14, Cobb 20°-40°, and Risser sign 0-2) were recruited from two clinics and randomly assigned to the PO and CO groups. Compliance sensors were embedded in both groups, while the PO group was set to adjust the interfacial pressure as prescribed automatically. Clinical assessments (radiology and quality of life, QoL) were conducted at the baseline, immediate after orthosis fitting and 1-year follow-up. Orthosis wearing compliance was tracked using thermo and pressure sensors. RESULTS: Twenty-four patients were enrolled with one drop-out (PO, n = 11; CO, n = 12). Significant immediate in-orthosis correction was observed in the PO (11.0° ± 6.5°, 42.0%, P < 0.001) and CO (10.3° ± 5.3°, 37.6%, P < 0.001) groups. After 1 year, no patient with PO progressed while two with CO had progression more than 5°. The mean daily wearing duration was 1.1 hours longer in the PO group as compared with the CO group (15.4 ± 4.5 vs. 14.3 ± 3.8 h). Moreover, the wearing quality within the targeted pressure was 33.9% higher in the PO group (56.5 ± 16.5% vs. 23.1 ± 12.1%, P < 0.001). No significant difference in the QoL results was observed between two groups nor within both groups during the study period. CONCLUSION: This study showed that the automated PO could enhance wearing quality when compared with the CO, thus offering a better biomechanical corrective effect in the study period without adverse effect on the patients' wearing quantity and QoL. LEVEL OF EVIDENCE: 1.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。

引用

  1. Lin Y, Lou E, Lam TP, et al. The Intelligent Automated Pressure-Adjustable Orthosis for Patients With Adolescent Idiopathic Scoliosis: A Bi-Center Randomized Controlled Trial. Spine (Phila Pa 1976). 2020. PMID: 32453223. doi:10.1097/BRS.0000000000003559

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。