文献速递关节外科

膝关节置换假体检索分析:松弛度与磨损的相关性

Retrieval Analysis of Cruciate-Retaining and Posterior-Stabilized Total Knee Arthroplasty and Correlations to Laxity and Wear.
PubMed 文献 · J Arthroplasty · 2020-03-21

该研究对47例初次全膝关节置换的尸体标本进行松弛度测量,并用半定量法评估胫骨垫磨损。结果发现后稳定型假体在60°和90°屈曲时内翻松弛度及60°屈曲时总冠状面松弛度显著大于后交叉韧带保留型,且后稳定型冠状面松弛度增加与胫骨磨损增加显著相关。

要点

  • 后稳定型假体在屈曲60°和90°时内翻松弛度显著大于后交叉韧带保留型。
  • 后稳定型假体冠状面松弛度增加与胫骨磨损增加显著相关,后交叉韧带保留型未见此趋势。
  • 无论使用何种假体,术中应避免造成过大屈曲松弛以优化长期聚乙烯磨损。

核心结论

  • 初次全膝关节置换后屈曲松弛度增大可能增加长期磨损,后稳定型假体因后交叉韧带对屈曲间隙支撑丧失可能更易受影响。

研究设计与样本

  • 对47例初次全膝关节置换的尸体标本在定制膝关节测试机上于完全伸直及30°、60°、90°屈曲位测量松弛度模式。
  • 采用基于Hood等1983年方法的半定量法评估胫骨垫磨损模式。

主要发现

  • 后稳定型假体队列在60°和90°屈曲时内翻松弛度较后交叉韧带保留型队列有统计学显著增加。
  • 后稳定型假体队列在60°屈曲时总冠状面松弛度较后交叉韧带保留型队列有统计学显著增加。
  • 后稳定型假体冠状面松弛度增加与胫骨磨损增加显著相关,后交叉韧带保留型标本未见此趋势。

临床怎么用

  • 无论使用后交叉韧带保留型还是后稳定型假体,术中应避免造成过大屈曲松弛以优化长期聚乙烯磨损。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Total knee arthroplasty (TKA) with posterior-stabilized (PS) or posterior cruciate-retaining (CR) implants has high success rates and survivorship. However, it is uncertain how laxity and constraint are associated with long-term polyethylene wear under physiological conditions. METHODS: To answer this question, we measured the laxity patterns of 47 harvested cadaver specimens with primary TKAs in a custom knee-testing machine at full extension and at 30°, 60°, and 90° of flexion. The wear patterns of the tibial inserts were assessed using a semiquantitative method which is a modified approach of that proposed by Hood et al in 1983. RESULTS: Statistical analysis found that the PS TKA cohort had a statistically significant increase in varus laxity at 60° and 90° of flexion, as well as total coronal laxity at 60° of flexion when compared to the CR cohort. Furthermore, analysis demonstrated a significant correlation between increased PS coronal laxity and increased tibial wear, a trend that was not seen in the CR specimens. CONCLUSION: Our findings suggest that greater laxity in flexion after primary TKA may increase the wear realized over time and that PS TKAs may be more susceptible due to the loss of support the PCL affords to the flexion space. Whether a CR or PS TKA is used, surgeons need to avoid the pitfalls that may create greater flexion laxity during the procedure to optimize long-term polyethylene wear.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Perkins MR, Arnholt CM, MacDonald DW, et al. Retrieval Analysis of Cruciate-Retaining and Posterior-Stabilized Total Knee Arthroplasty and Correlations to Laxity and Wear. J Arthroplasty. 2020. PMID: 32279944. doi:10.1016/j.arth.2020.03.027

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。