回收髋臼杯多孔涂层骨长入评估
本研究对回收的钛纤维网(HG)与钽金属(TM)涂层髋臼杯进行骨体积分数评估,按年龄、BMI和性别匹配8对。HG组平均骨体积分数21±17%,TM组7±4%(P<.0001);边缘和极点区骨长入少于赤道区,两组骨长入与植入时间均无关联。
要点
- HG涂层髋臼杯骨体积分数(21±17%)显著高于TM涂层(7±4%)。
- 边缘和极点区骨长入均少于赤道区。
- 两种设计骨长入与植入时间均无关联。
核心结论
- 充足的骨长入是成功生物固定的要求,但骨长入量可能不是驱动因素。
- 两种植入物尽管骨长入量低,仍获得了成功结局和长期固定。
研究设计与样本
- 回收髋臼杯对照研究,8个HG壳体与8个TM壳体按患者年龄、BMI和性别匹配。
- 平均手术年龄69(51-82)岁,平均BMI 28(21-40),男女各4例。
- TM组植入时间40(16-96)个月,HG组156(108-216)个月。
- 每个部件经顶点切取两个5-mm厚横截面,行背散射扫描电子显微镜评估。
主要发现
- HG壳体平均骨体积分数为21±17%,TM壳体为7±4%(P < .0001)。
- 边缘和极点区域骨长入均少于赤道区域。
- 两种设计的骨长入与植入时间均无关联。
临床怎么用
- 骨长入量低不必然导致生物固定失败,评估髋臼杯固定时应结合长期临床结局。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: We evaluated bone volume fraction in retrieved acetabular shells with 2 types of porous coatings: (1) titanium fiber mesh (HG) components and (2) tantalum metal coating (TM) components. METHODS: Eight HG shells were matched to 8 TM shells for patient age, body mass index, and gender. The mean age at index surgery was 69 (51-82) years, mean body mass index was 28 (21-40), and patients were evenly divided between male and female (4:4). The length of implantation was 40 (16-96) months for the TM group and 156 (108-216) months for the HG group. Shells were embedded and two 5-mm thick cross-sections were cut through the apex of each component for backscatter scanning electron microscopy assessment. Backscatter scanning electron microscopy images were segmented to threshold for metal, bone, and available space for ingrowth. Slices were assessed regionally for ingrowth at the rim, equator, and pole of the acetabular shell. Differences were assessed using general estimating equations, and P values were adjusted for multiple comparisons using the Holm-Bonferroni step-down procedure. RESULTS: The mean bone volume fraction was 21 ± 17% for the HG shell and 7 ± 4% for the TM shell (P < .0001). The rim and pole regions both had less bone ingrowth than the equator. No association was found between bone ingrowth and length of implantation for either design. CONCLUSION: Adequate bone ingrowth is a requirement for successful biological fixation, but the amount of ingrowth may not be a driving factor. Both implants studied had successful outcomes and long-term fixation despite the observation of low amounts of ingrowth.
引用
- Baral EC, Trivellas M, Vigdorchik JM, et al. Porous Coatings in Retrieved Acetabular Components. J Arthroplasty. 2020. PMID: 32307292. doi:10.1016/j.arth.2020.03.036