前路椎体生长调控:患者特异性有限元规划工具2年预测能力及生长调控生物力学评估
结果显示FEM对术后即刻、1年和2年冠状面三维矫正指数预测误差在4°以内,脊柱高度误差<0.8 cm(±2%);即刻矫正40%,1年和2年分别额外矫正13%和3%。
要点
- FEM规划工具对AVBGM术后2年冠状面矫正预测误差≤4°,脊柱高度误差<0.8 cm。
- AVBGM术后即刻矫正约40%,1年额外矫正13%,2年额外矫正3%。
- AVBGM安装后顶椎区生长板凸凹侧计算力差降低39%,椎间盘降低46%。
核心结论
- FEM规划工具可提供有临床相关性的矫正预测,有助于合理化手术规划。
- AVBGM对生长调控的生物力学效应在术后第一年内似乎达到最大。
研究设计与样本
- 利用FEM评估不同器械配置,并选择预测2年矫正最优的策略用于手术。
- 计算模拟的三维矫正指数及椎体生长板、椎间盘和器械上的应力。
主要发现
- 平均每例测试6种器械配置。
- 术后即刻、1年和2年冠状面三维矫正指数预测值与实际结果差异在4°以内,脊柱高度差异<0.8 cm(±2%)。
- 术后即刻矫正为40%,术后1年和2年分别额外矫正13%和3%。
- AVBGM安装后顶椎水平生长板凸凹侧计算力差降低39%,椎间盘降低46%。
临床怎么用
- 可利用患者特异性FEM在术前模拟不同AVBGM器械配置,辅助选择预测2年矫正最优的方案。
- 应关注AVBGM术后第一年的生长调控生物力学效应,该阶段效应似乎最大。
局限性
- 证据等级为3级。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
STUDY DESIGN: Numerical planning and simulation of immediate and after 2 years growth modulation effects of anterior vertebral body growth modulation (AVBGM). OBJECTIVE: The objective was to evaluate the planning tool predictive capability for immediate, 1-year, and 2-year postoperative correction and biomechanical effect on growth modulation over time. SUMMARY OF BACKGROUND DATA: AVBGM is used to treat pediatric scoliotic patients with remaining growth potential. A planning tool based on a finite element model (FEM) of pediatric scoliosis integrating growth was previously developed to simulate AVBGM installation and growth modulation effect. METHODS: Forty-five patients to be instrumented with AVBGM were recruited. A patient-specific FEM was preoperatively generated using a 3D reconstruction obtained from biplanar radiographs. The FEM was used to assess different instrumentation configurations. The strategy offering the optimal 2-year postoperative correction was selected for surgery. Simulated 3D correction indices, as well as stresses applied on vertebral epiphyseal growth plates, intervertebral discs, and instrumentation, were computed. RESULTS: On average, six configurations per case were tested. Immediate, 1-year, and 2-year postoperative 3D correction indices were predicted within 4° of that of actual results in coronal plane, whereas it was <0.8 cm (±2%) for spinal height. Immediate postoperative correction was of 40%, whereas an additional correction of respectively 13% and 3% occurred at 1- and 2 year postoperative. The convex/concave side computed forces difference at the apical level following AVBGM installation was decreased by 39% on growth plates and 46% on intervertebral discs. CONCLUSION: This study demonstrates the FEM clinical usefulness to rationalize surgical planning by providing clinically relevant correction predictions. The AVBGM biomechanical effect on growth modulation over time seemed to be maximized during the first year following the installation. LEVEL OF EVIDENCE: 3.
引用
- Cobetto N, Aubin CE, Parent S, et al. Anterior Vertebral Body Growth Modulation: Assessment of the 2-year Predictive Capability of a Patient-specific Finite-element Planning Tool and of the Growth Modulation Biomechanics. Spine (Phila Pa 1976). 2020. PMID: 32341305. doi:10.1097/BRS.0000000000003533
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。