文献速递创伤骨科骨质疏松与代谢性骨病

螺钉位置是否影响股骨颈固定后粗隆下骨折风险?一项生物力学研究

Does Screw Location Affect the Risk of Subtrochanteric Femur Fracture After Femoral Neck Fixation? A Biomechanical Study.
PubMed 文献 · Clin Orthop Relat Res · 2020-01-01

该生物力学研究在正常密度和骨质疏松Sawbones股骨上,比较最远端螺钉起点位于小转子近端、平齐或远端时对股骨颈固定后粗隆下骨折风险和失效载荷的影响。结果显示在骨质疏松标本中,螺钉起点位于小转子远端时粗隆下骨折比例更高且失效载荷更低。

要点

  • 骨质疏松标本中,最远端螺钉起点位于小转子远端时粗隆下骨折比例更高。
  • 最远端螺钉起点位于小转子远端时,正常密度和骨质疏松标本的失效载荷均更低。
  • 避免最远端螺钉起点位于小转子远端或可降低骨质疏松患者粗隆下骨折风险。

核心结论

  • 在骨质疏松标本中,最远端螺钉起点位于小转子远端与粗隆下股骨骨折相关,且失效载荷降低。

研究设计与样本

  • 使用正常密度和骨质疏松Sawbones股骨,以三角形顶点远端配置置入三枚空心螺钉,最远端螺钉起点分别位于小转子近端、平齐或远端,沿机械轴加载至失效。

主要发现

  • 骨质疏松标本中,螺钉起点位于小转子远端时粗隆下骨折比例为3/10,而起点位于小转子平齐或近端时为0/20(p=0.030)。
  • 正常密度标本中未观察到粗隆下骨折。
  • 最远端螺钉起点位于小转子远端时失效载荷更低:正常密度亚组为13,502±1980 N vs 14,675±1528 N;骨质疏松亚组为8946±1509 N vs 10,026±1256 N;线性回归系数1127 N(95%CI 298至1956 N;调整r=0.71;p=0.009)。

关键数值与推荐

  • 骨质疏松标本中,最远端螺钉起点位于小转子远端时粗隆下骨折比例为3/10,起点位于小转子平齐或近端时为0/20(p=0.030)。
  • 最远端螺钉起点位于小转子远端时失效载荷更低:正常密度亚组13,502±1980 N vs 14,675±1528 N;骨质疏松亚组8946±1509 N vs 10,026±1256 N;线性回归系数1127 N(95%CI 298至1956 N;调整r=0.71;p=0.009)。

临床怎么用

  • 在股骨颈骨折固定中,避免最远端螺钉起点位于小转子水平以远,可能降低灾难性粗隆下股骨骨折风险,尤其对骨质疏松患者。
  • 更整体的内翻螺钉排列理论上可能影响骨折端加压,进而对急性期骨折愈合产生影响。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Case reports suggest that there is an increased risk of subtrochanteric femur fracture after femoral neck fixation with cannulated screws when the distal-most screw is placed distal to the lesser trochanter. However, to our knowledge, there are no biomechanical data supporting this observation. QUESTIONS/PURPOSES: (1) Is there an increased risk of subtrochanteric femur fracture after femoral neck fixation with cannulated screws in normal density and osteoporotic Sawbones when the distal-most screw is started distal to the lesser trochanter? (2) Does the screw starting point position after femoral neck fixation with cannulated screws affect load to failure when normal density and osteoporotic Sawbones are loaded through their mechanical axis? METHODS: Normal density and osteoporotic Sawbones femora were instrumented with three cannulated screws in a triangular apex distal configuration with the distal-most screw starting either proximal to, at, or distal to the level of the lesser trochanter. Specimens were loaded along the mechanical axis to failure. The fracture location and ultimate load to failure were compared between groups. RESULTS: The screw start point distal to the lesser trochanter resulted in a greater proportion of subtrochanteric femur fractures compared with screw start points at or proximal to the lesser trochanter in the subset of osteoporotic specimens (three of 10 specimens versus 0 of 20 specimens; p = 0.030). No subtrochanteric femur fractures were observed in the normal density specimens. Load to failure was lower when the distal-most screw was started distal to the lesser trochanter than when it was started at or proximal to the lesser trochanter (normal density subset 13,502 ± 1980 N versus 14,675 ±1528 N; osteoporotic subset 8946 ± 1509 N versus 10,026 ± 1256 N; linear regression coefficient 1127 N [95% CI 298 to 1956 N]; adjusted r = 0.71; p = 0.009). CONCLUSIONS: A screw start point distal to the lesser trochanter was associated with subtrochanteric femur fractures in the osteoporotic subset. Additionally, there was decreased load to failure when the distal-most screw was started distal to the lesser trochanter. CLINICAL RELEVANCE: These data suggest that avoiding a screw start point distal to the level of the lesser trochanter in femoral neck fracture fixation may decrease the risk of catastrophic subtrochanteric femur fractures, especially in patients with osteoporosis. However, it should be noted that a more overall varus screw alignment could theoretically compromise the ability to achieve compression across the fracture, with attendant implications with regard to fracture union in the acute setting.

来源:Clin Orthop Relat Res,查看原文。版权归原作者所有。

引用

  1. Crump EK, Quacinella M, Deafenbaugh BK, et al. Does Screw Location Affect the Risk of Subtrochanteric Femur Fracture After Femoral Neck Fixation? A Biomechanical Study. Clin Orthop Relat Res. 2020. PMID: 32229749. doi:10.1097/CORR.0000000000000945

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。