模块化双动全髋翻修术后血清金属离子升高风险
结论提示MDM翻修术后可能因腐蚀产生金属碎屑导致血清金属离子升高。
要点
- 6例(16.2%)离子水平超过5 μg/L,5例(13.5%)在1-5 μg/L之间。
核心结论
- MDM全髋翻修术后血清金属离子升高可能继发于假体腐蚀产生的金属碎屑。
- 在年轻活跃患者的翻修关节置换决策中应考虑这一潜在风险。
研究设计与样本
- 所有患者均接受小梁钛MDM髋臼组件,采用质谱法测量血清钴和铬水平。
主要发现
- 血清铬平均2.08 μg/L(95%CI 0.9-3.2,范围0.02-11.8)。
- 血清钴平均1.99 μg/L(95%CI 0.81-3.17,范围0.07-16.05)。
关键数值与推荐
- 血清铬平均2.08 μg/L(95%CI 0.9-3.2),钴平均1.99 μg/L(95%CI 0.81-3.17)。
临床怎么用
- 对接受MDM全髋翻修术的患者,尤其是年轻活跃者,应监测血清金属离子水平。
- 在翻修关节置换决策中应将MDM组件腐蚀导致金属离子升高的潜在风险纳入考量。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Modular dual-mobility (MDM) total hip arthroplasty (THA) is designed with a cobalt-chromium liner inserted into a titanium acetabular component. The purpose of this study is to investigate the potential risks for fretting corrosion at this junction, by measuring serum metal ions, after MDM acetabular revision. METHODS: Thirty-seven patients with well-functioning revision THAs participated in a cross-sectional study at mean 5.1 (2-10) years after surgery. All received a trabecular titanium MDM acetabular component. The serum levels of cobalt and chromium were measured using mass spectrometry. RESULTS: The mean values of chromium and cobalt were 2.08 μg/L (95% confidence interval 0.9-3.2, range 0.02-11.8) and 1.99 μg/L (95% confidence interval 0.81-3.17, range 0.07-16.05), respectively. Eleven patients (29, 7%) had ion levels above the normal range, with 6 (16.2%) above 5 μg/L and 5 (13.5%) between 1 and 5 μg/L. A significant correlation was found between an elevated serum metal ion level and University of California Los Angeles score (P = .016). CONCLUSION: We conclude that serum metal level elevation may occur secondary to metal debris resulting from corrosion of the index MDM THA. This potential risk should be included in the decision-making process when dealing with revision arthroplasty in young and active patients.
引用
- Civinini R, Cozzi Lepri A, Carulli C, et al. Patients Following Revision Total Hip Arthroplasty With Modular Dual Mobility Components and Cobalt-Chromium Inner Metal Head are at Risk of Increased Serum Metal Ion Levels. J Arthroplasty. 2020. PMID: 32197964. doi:10.1016/j.arth.2020.02.050
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。