全髋与全膝关节置换术后关节液自动白细胞计数假阳性风险需警惕
该回顾性研究在同一实验室纳入8607例关节置换术后和812例原生膝关节的关节液样本,每例自动白细胞计数>3000 cells/μL且配有反射手工计数。自动与手工计数相关性在原生膝接近完美、全膝置换强、全髋置换中等;假阳性率分别为4.4%、10.1%和34.3%(P < .0001)。结论提示关节置换显著增加自动白细胞计数假阳性风险,临床解读时应考虑反射手工计数验证。
要点
- 全髋置换关节液自动白细胞计数假阳性率高达34.3%,全膝为10.1%,原生膝为4.4%(P < .0001)。
- 自动与手工计数相关性:原生膝接近完美、全膝置换强、全髋置换中等。
- 国际共识会议评分和培养阳性证实手工计数而非自动计数正确。
核心结论
- 髋或膝关节置换的存在似乎大幅增加关节液自动白细胞计数假阳性风险。
- 临床评估关节置换患者时,解读自动白细胞计数阳性应谨慎,并考虑请求反射手工计数以验证准确性。
研究设计与样本
- 单实验室回顾性综述,识别8607例连续关节置换术后关节液样本和812例原生膝关节样本,每例自动白细胞计数>3000 cells/μL且配有配对反射手工计数。
- 评估自动与手工计数的相关性及自动白细胞计数假阳性率。
主要发现
- 原生膝、全膝置换和全髋置换的自动白细胞计数相关性分别为接近完美、强和中等。
- 自动计数假阳性率分别为原生膝4.4%、全膝置换10.1%、全髋置换34.3%(P < .0001)。
- 国际共识会议评分和培养阳性表明手工计数而非自动计数正确。
关键数值与推荐
- 自动白细胞计数>3000 cells/μL的样本纳入并配对手工计数。
- 假阳性率:原生膝4.4%、全膝置换10.1%、全髋置换34.3%(P < .0001)。
临床怎么用
- 评估关节置换患者关节液时,对自动白细胞计数阳性结果应谨慎解读。
- 考虑请求反射手工计数以验证自动细胞计数的准确性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Although false elevation of automated leukocyte (white blood cell [WBC]) counts has been described in the setting of hip corrosion, the more general correlation between manual and automated cell counts among synovial fluid aspirates from hip and knee arthroplasties has not been studied. METHODS: This retrospective review at one laboratory identified 8607 consecutive synovial fluid samples from arthroplasties and 812 from native knees, each with an automated WBC count > 3000 cells/μL and a corresponding paired reflex manual count. The correlation between automated and manual counts was evaluated, as was the rate of false-positive automated WBC counts. RESULTS: The correlation between automated WBC counts for native knees, total knee arthroplasties, and total hip arthroplasties was near-perfect, strong, and moderate, respectively. The false-positive rates for automated counts were 4.4%, 10.1%, and 34.3%, respectively (P < .0001). International Consensus Meeting scores and culture positivity demonstrated that manual counts, not automated counts, were correct. CONCLUSION: The presence of a hip or knee arthroplasty appears to substantially increase the risk of a false-positive automated synovial fluid WBC count. Clinicians evaluating an arthroplasty should exercise caution when interpreting positive automated WBC counts, and consider requesting a reflex manual count, to verify the accuracy of automated cell counting.
引用
- Deirmengian CA, Kazarian GS, Feeley SP, et al. False-Positive Automated Synovial Fluid White Blood Cell Counting Is a Concern for Both Hip and Knee Arthroplasty Aspirates. J Arthroplasty. 2020. PMID: 32113809. doi:10.1016/j.arth.2020.01.060
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。