全膝关节置换术后胫骨无菌性松动与骨水泥层质量相关
本研究基于机构假体回收数据库,纳入2007至2015年间因胫骨无菌性松动行翻修的全膝关节置换术(TKA)61例,并与59例至少3年未失败TKA匹配对照。术后6周X线片分析显示,骨水泥渗透深度不足2mm的区域数、假体-骨水泥界面透亮线百分比及胫骨假体内翻角度与无菌性松动独立相关。
要点
- 骨水泥渗透不足2mm的区域数每增加1个,无菌性松动风险增加(OR 1.89)。
- 假体-骨水泥界面透亮线百分比每增加1%,失败风险增加(OR 1.15)。
- 胫骨假体内翻角度每增加1度,失败风险增加(OR 1.35)。
核心结论
- 骨水泥层质量差的影像学指标与后期胫骨无菌性松动相关。
研究设计与样本
- 病例对照研究,纳入61例因胫骨无菌性松动行翻修的TKA及59例至少3年未失败的匹配对照TKA。
主要发现
- 失败组骨水泥渗透<2mm的区域数多于对照组(5.6 vs 3.4),OR 1.89,P < .001。
- 失败组假体-骨水泥界面透亮线百分比高于对照组(8.7% vs 3.1%),OR 1.15,P = .001。
- 失败组胫骨假体内翻角度大于对照组(1.5° vs 0°),OR 1.35,P = .014。
- 骨-骨水泥界面透亮线区域数在两组间无显著差异(1.1 vs 0.3,P = .091)。
临床怎么用
- 术中应重视骨水泥技术,确保充分的骨水泥渗透并避免胫骨假体内翻,以降低TKA失败风险。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Aseptic loosening remains one of the leading causes for failure of total knee arthroplasty (TKA). We sought to identify early radiographic measures that may associate with aseptic tibial component loosening, emphasizing systematic evaluation of the cement mantle. METHODS: All TKA revisions from 2007 to 2015 with the primary indication of tibial aseptic loosening were identified using in an institutional implant retrieval database. After exclusion criteria, 61 TKAs comprised the study group. A matched control group of 59 TKAs that had not failed at a minimum of 3 years was identified for comparison. Radiographic analysis on all 6-week postoperative radiographs included angulation of components, cement penetration depth, and presence of radiolucency at the implant-cement and bone-cement interfaces. Groups were compared with Student's t-test, chi-squared test, and Mann-Whitney U-test. A final multivariable logistic regression model was formed for the outcome of aseptic loosening. RESULTS: On multivariable analysis, failure was associated with a greater number of zones with cement penetration <2 mm (5.6 vs 3.4 zones, odds ratio [OR] 1.89, P < .001), increasing percent involvement of radiolucency at the implant-cement interface (8.7% vs 3.1%, OR = 1.15, P = .001), and increased varus alignment of the tibial component (1.5° vs 0°, OR = 1.35, P = .014). A greater number of zones with a radiolucent line at the bone-cement interface did not significantly associate (1.1 vs 0.3, P = .091). CONCLUSION: Our results suggest that radiographic indicators of poor cement mantle quality associate with later aseptic loosening. This emphasizes the need for surgeons to perform careful cement technique in order to reduce the risk of TKA failure. LEVEL OF EVIDENCE: III (Case-control).
引用
- Hampton CB, Berliner ZP, Nguyen JT, et al. Aseptic Loosening at the Tibia in Total Knee Arthroplasty: A Function of Cement Mantle Quality? J Arthroplasty. 2020. PMID: 32171492. doi:10.1016/j.arth.2020.02.028
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。