模块化颈全髋关节置换患者五年随访研究
该研究回顾性分析186例模块化颈全髋关节置换患者,平均随访60.1个月,发现因颈-柄腐蚀翻修率达22.0%,生存率3年为87%、7年为72%,中期翻修率较早期报道几乎翻倍。
要点
- 模块化颈全髋因颈-柄腐蚀中期翻修率22.0%,生存率3年87%、7年72%。
- 翻修患者均有腹股沟痛或局部肿胀症状,症状是失败的关键特征。
- 血清钴离子水平随时间升高,但仅凭症状随访即可识别失败,无需常规血检。
核心结论
- 模块化颈全髋关节置换因颈-柄腐蚀的中期翻修率较短期结果几乎翻倍。
研究设计与样本
- 回顾性连续队列研究,纳入175例患者共186例模块化颈髋关节,平均随访60.1±22.9个月(范围24-100)。
主要发现
- 因颈-柄腐蚀翻修41髋(22.0%)。
- 翻修患者临床均有腹股沟痛伴或不伴局部肿胀。
- 所有患者血清钴离子水平随时间升高。
- 影像学显示有症状患者多见较大液体积聚和局部软组织肿块。
- 14%翻修患者MRI或超声无阳性发现。
关键数值与推荐
- 生存率3年为87%(95%CI 81-92),7年为72%(95%CI 64-80)。
临床怎么用
- 临床随访症状(腹股沟痛±局部肿胀)是识别模块化颈失败的主要依据,可考虑单独临床随访而非常规血离子检测。
局限性
- 回顾性研究设计。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: We have previously reported the early clinical results of a modular-neck stem identifying an early 2-year revision rate of 13% due to neck-stem corrosion. This report updates our findings to a midterm mean follow-up of 5 years. METHODS: This is a consecutive retrospective review of 186 modular-neck hips in 175 patients with a mean follow-up period was 60.1 ± 22.9 months (range 24-100). We reviewed clinical findings, routine radiographs, detailed imaging (metal artifact reduction software-magnetic resonance imaging, ultrasound), and serum ion levels of cobalt and chromium. We performed a survival analysis with the endpoint defined as revision total hip arthroplasty due to neck-stem corrosion. RESULTS: We revised 41 hips (22.0%) for neck-stem corrosion. Clinical symptoms (groin pain ± local swelling) were consistently present in those that came for revision. Mean serum cobalt ion levels increased as time passed in all patients. Detailed image findings showed that larger fluid collections and local soft tissue masses were seen predominately in symptomatic patients. However, 14% of patients who underwent revision did not have positive magnetic resonance imaging or ultrasound findings. The survival rate was 87% (95% confidence interval 81-92) at 3 years postoperatively and 72% (95% confidence interval 64-80) at 7 years postoperatively. CONCLUSION: The revision rate for this modular-neck stem due to neck-stem corrosion at mid-term follow-up almost doubled in comparison to previous short-term results. It seems reasonable to consider clinical follow-up alone as symptoms, rather than blood testing for ion levels, seem to be the defining characteristic of failure.
引用
- Inoue D, Restrepo C, Nourie B, et al. Patients With Modular-Neck Total Hip Arthroplasty: A Brief Five-Year Follow-Up Study. J Arthroplasty. 2020. PMID: 32199762. doi:10.1016/j.arth.2020.02.059
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。