不同骨缺损尺寸对非骨水泥髋臼杯固定稳定性的影响

Fixation Stability of Uncemented Acetabular Cups With Respect to Different Bone Defect Sizes.
PubMed 文献 · J Arthroplasty · 2020-01-18

本研究通过生物力学模型评估非骨水泥髋臼压配杯在不同标准化髋臼骨缺损下的固定稳定性,发现内外侧覆盖缺失50%为临界值,20%至30%缺失可接受,50%/50%缺损为需附加固定的阈值。

要点

  • 内外侧覆盖缺失50%被视为非骨水泥杯固定的临界值。
  • 内外侧覆盖缺失20%至30%在存在初始稳定性时可接受。
  • 50%/50%缺损被确定为使用附加固定方法的阈值。

核心结论

  • 内外侧髋臼覆盖缺失50%被评估为非骨水泥杯固定的临界值,此时植入物与骨在缺损区的接触带丧失。
  • 内外侧覆盖缺失20%至30%在存在初始稳定性时对压配杯固定似乎可接受。
  • 50%/50%缺损被确定为使用附加固定方法的阈值。

研究设计与样本

  • 开发髋臼缺损模型并转移至生物力学杯-块模型,比较11种不同缺损腔与完整腔在压入和撬出力测试中的表现,使用一种压配杯设计(外径56 mm)。

主要发现

  • 撬出力矩范围从15.53 ± 1.38 Nm(完整腔)至1.37 ± 0.54 Nm(83%/50%缺损)。
  • 较小缺损(33%/11%、33%/22%和50%/11%)使撬出力矩平均降低33.9% ± 2.8%。

关键数值与推荐

  • 内外侧覆盖缺失50%为临界值。
  • 内外侧覆盖缺失20%至30%可接受。
  • 50%/50%缺损为使用附加固定方法的阈值。

临床怎么用

  • 对于内外侧覆盖缺失20%至30%的髋臼骨缺损,若存在初始稳定性可考虑使用压配杯固定。
  • 内外侧覆盖缺失达50%时需谨慎评估固定稳定性。
  • 50%/50%缺损时应考虑使用附加固定方法。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: In total hip arthroplasty, acetabular press-fit cups require a proper bone stock for sufficient primary implant fixation. The presence of acetabular bone defects compromises the primary fixation stability of acetabular press-fit cups. The aim of the present study is to determine the fixation stability of a cementless acetabular cup regarding standardized bone defects in an experimental setup. METHODS: An acetabular defect model was developed and transferred to a biomechanical cup-block model. The lack of superior cup coverage was divided into 4 stages of superior rim loss (33%, 50%, 67%, and 83%) in the anterior-posterior direction and into 4 stages of mediolateral wall absence (11%, 22%, 33%, and 50%). This resulted in 11 different defect cavities, which were compared to the intact cavity in push-in and lever-out tests of one press-fit cup design (56 mm outer diameter). Thereby, push-in force, lever-out moment, lever-out angle, and interface stiffness were determined. RESULTS: The determined lever-out moments range from 15.53 ± 1.38 Nm (intact cavity) to 1.37 ± 0.54 Nm (83%/50% defect). Smaller defects (33%/11%, 33%/22%, and 50%/11%) reduce the lever-out moments by an average of 33.9% ± 2.8%. CONCLUSION: The lack of mediolateral acetabular coverage of 50% was assessed as critical for cementless cup fixation, whereby the contact zone between implant and bone in the defect is lost. A lack of 20% to 30% mediolateral coverage appears to be acceptable for press-fit cup fixation in the presence of primary stability. A defect of 50%/50% was identified as the threshold for using additional fixation methods.