增强现实辅助直接前入路全髋关节置换髋臼假体放置
该研究在射线可透泡沫骨盆模型上模拟髋臼假体放置,比较新型三维增强现实(AR)导航系统与标准透视引导的准确性、精确性、手术时间、辐射剂量和可用性。
要点
- AR导航在泡沫骨盆模型上提高髋臼假体外倾角和前倾角的准确性及前倾角精确性。
核心结论
- 新型三维AR导航系统在髋臼假体放置中产生更准确的外倾角和前倾角以及更精确的前倾角。
- AR导航比标准透视更快、更易使用,且不增加辐射剂量。
研究设计与样本
- 在射线可透泡沫骨盆模型上模拟髋臼假体放置。
- 将锥形束计算机断层扫描数据与红绿蓝深度相机的光学数据配准以创建AR环境。
- 八名骨科手术培训生使用两种方法完成假体放置。
临床怎么用
- AR导航可辅助培训生更准确、更快速地放置髋臼假体,且不增加辐射暴露。
- 结果基于泡沫骨盆模型和培训生操作,临床转化需进一步验证。
局限性
- 研究在射线可透泡沫骨盆模型上进行模拟,非真实患者。
- 仅纳入八名骨科手术培训生,样本量小。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Malposition of the acetabular component of a hip prosthesis can lead to poor outcomes. Traditional placement with fluoroscopic guidance results in a 35% malpositioning rate. We compared the (1) accuracy and precision of component placement, (2) procedure time, (3) radiation dose, and (4) usability of a novel 3-dimensional augmented reality (AR) guidance system vs standard fluoroscopic guidance for acetabular component placement. METHODS: We simulated component placement using a radiopaque foam pelvis. Cone-beam computed tomographic data and optical data from a red-green-blue-depth camera were coregistered to create the AR environment. Eight orthopedic surgery trainees completed component placement using both methods. We measured component position (inclination, anteversion), procedure time, radiation dose, and usability (System Usability Scale score, Surgical Task Load Index value). Alpha = .05. RESULTS: Compared with fluoroscopic technique, AR technique was significantly more accurate for achieving target inclination (P = .01) and anteversion (P = .02) and more precise for achieving target anteversion (P < .01). AR technique was faster (mean ± standard deviation, 1.8 ± 0.25 vs 3.9 ± 1.6 minute; P < .01), and participants rated it as significantly easier to use according to both scales (P < .05). Radiation dose was not significantly different between techniques (P = .48). CONCLUSION: A novel 3-dimensional AR guidance system produced more accurate inclination and anteversion and more precise anteversion in the placement of the acetabular component of a hip prosthesis. AR guidance was faster and easier to use than standard fluoroscopic guidance and did not involve greater radiation dose.
引用
- Alexander C, Loeb AE, Fotouhi J, et al. Augmented Reality for Acetabular Component Placement in Direct Anterior Total Hip Arthroplasty. J Arthroplasty. 2020. PMID: 32063415. doi:10.1016/j.arth.2020.01.025
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。