标准与高边高交联聚乙烯髋臼内衬在初次全髋关节置换术中的手术结局比较:至少15年随访的单中心回顾性队列研究
该单中心回顾性队列研究纳入2000年3月至2003年12月由同一资深关节置换外科医生完成的112例初次全髋关节置换术,按髋臼内衬类型分为标准组和高边组。结果显示两组线性和体积磨损率无显著差异,标准组确认1例骨溶解而高边组无骨溶解。作者认为高边高交联聚乙烯髋臼内衬可能是可安全使用的良好植入物选择。
要点
- 标准组与高边组线性及体积磨损率均无显著差异。
- 标准组确认1例骨溶解,高边组无骨溶解。
- 高边高交联聚乙烯髋臼内衬可能是可安全使用的植入物选择。
核心结论
- 高边高交联聚乙烯髋臼内衬可能是可安全使用的良好植入物选择。
研究设计与样本
- 单中心回顾性队列研究,评估2000年3月至2003年12月由同一资深关节置换外科医生完成的112例初次全髋关节置换术。
- 患者按所用髋臼内衬类型分为标准内衬组和高边内衬组。
主要发现
- 标准组线性磨损率为0.028 mm/y(0.000-0.145 mm/y),体积磨损率为11.641 mm3/y(0.000-70.000 mm3/y)。
- 高边组线性磨损率为0.026 mm/y(0.000-0.094 mm/y),体积磨损率为9.706 mm3/y(0.000-33.000 mm3/y)。
- 两组磨损率无显著差异。
- 标准组确认1例骨溶解,高边组无骨溶解。
临床怎么用
- 在初次全髋关节置换术中可考虑使用高边高交联聚乙烯髋臼内衬,其至少15年随访的磨损与骨溶解结局与标准内衬相当。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The aims of the present study are to (1) conduct the longest-to-date follow-up wear analysis of the highly cross-linked polyethylene acetabular liners, (2) assess the incidence of femoral and acetabular osteolysis, and (3) compare the surgical outcomes between standard and elevated-rim acetabular liners in primary total hip arthroplasty (THA). METHODS: In this retrospective cohort study, we evaluated 112 primary THAs performed by a single experienced arthroplasty surgeon at our institution between March 2000 and December 2003. Patients were classified based on the type of acetabular liner used: standard or elevated-rim liner. For evaluation of surgical outcomes, the following data were collected: acetabular cup position (anteversion and inclination), wear rate (linear and volumetric), presence of osteolysis, history of reoperation (all-cause and wear-related), complications (deep joint infection, dislocation, and periprosthetic fracture), and Harris hip score at last follow-up. RESULTS: Linear and volumetric wear rates were 0.028 mm/y (0.000-0.145 mm/y) and 11.641 mm3/y (0.000-70.000 mm3/y) in the standard group and 0.026 mm/y (0.000-0.094mm/y) and 9.706 mm3/y (0.000-33.000 mm3/y) in the elevated-rim group, respectively. These rates were not significantly different between groups. One case of osteolysis was confirmed in the standard group, whereas no osteolysis was observed in the elevated-rim group. CONCLUSION: We suggest that elevated-rim highly cross-linked polyethylene acetabular liners might be a good implant option that can be used safely.
引用
- Shin WC, Moon NH, Jeon SB, et al. Comparison of Surgical Outcomes Between Standard and Elevated-Rim Highly Cross-Linked Polyethylene Acetabular Liners in Primary Total Hip Arthroplasty With Minimum 15-Year Follow-Up: Single-Center, Retrospective Cohort Study. J Arthroplasty. 2019. PMID: 31983565. doi:10.1016/j.arth.2019.12.026