文献速递关节外科

聚乙烯制造特性对非骨水泥型与混合型全髋关节置换翻修风险有重大影响

Polyethylene manufacturing characteristics have a major effect on the risk of revision surgery in cementless and hybrid total hip arthroplasties.
PubMed 文献 · Bone Joint J · 2020-01-01

该研究基于国家关节登记与厂商数据的前瞻性队列,分析聚乙烯髋臼内衬制造特性对非骨水泥型和混合型全髋关节置换生存率的影响。主要终点为无菌性松动翻修,采用Cox回归分析。结果显示总辐射剂量、内衬面不对称性和稳定化处理显著影响生存率;总辐射剂量≥5 Mrad、不对称内衬面及熔点以上加热稳定化者翻修风险最低。

要点

  • 总辐射剂量≥5 Mrad(G3、G4)较无辐射G1的无菌性松动翻修风险显著降低(校正HR 0.36、0.38)。
  • 不对称内衬面较对称(平面)内衬的无菌性松动翻修风险更低。
  • G3和G4中,维生素E稳定化且加热至熔点以上者表现最佳。

核心结论

  • 聚乙烯内衬总辐射剂量≥5 Mrad、内衬面不对称且采用熔点以上加热稳定化时,植入物生存率最佳。

研究设计与样本

  • 前瞻性队列研究,链接英国国家关节登记(NJR)与厂商数据,主要终点为无菌性松动翻修,采用Cox比例风险回归分析。
  • 共记录5,329例全髋关节置换翻修,其中1,290例因无菌性松动翻修。

主要发现

  • 总辐射剂量、内衬面不对称性和稳定化处理显著影响植入物生存。
  • G1(无辐射)任何原因翻修及无菌性松动翻修风险最高,G3和G4最低。
  • 相比G1,G2的校正风险比为0.74(95%CI 0.64–0.86),G3为0.36(95%CI 0.30–0.43),G4为0.38(95%CI 0.31–0.47)。
  • 12年无菌性松动翻修的累积发生率:G3为0.52/100例THA,G4为0.54/100例THA,G1为1.95/100例THA。
  • 不对称内衬较对称(平面)内衬的无菌性松动翻修风险及非无菌性松动翻修风险均更低。

关键数值与推荐

  • 总辐射剂量分组:G1无辐射;G2>0 Mrad且<5 Mrad;G3≥5 Mrad且<10 Mrad;G4≥10 Mrad。
  • G3和G4中,维生素E稳定化且加热至熔点以上表现最佳。

临床怎么用

  • 选择非骨水泥型或混合型全髋关节置换的聚乙烯内衬时,优先考虑总辐射剂量≥5 Mrad的产品。
  • 优先选择内衬面不对称且采用熔点以上加热稳定化(如维生素E)的聚乙烯内衬,以降低翻修风险。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The aim of this study was to identify the effect of the manufacturing characteristics of polyethylene acetabular liners on the survival of cementless and hybrid total hip arthroplasty (THA). METHODS: Prospective cohort study using linked National Joint Registry (NJR) and manufacturer data. The primary endpoint was revision for aseptic loosening. Cox proportional hazard regression was the primary analytical approach. Manufacturing variables included resin type, crosslinking radiation dose, terminal sterilization method, terminal sterilization radiation dose, stabilization treatment, total radiation dose, packaging, and face asymmetry. Total radiation dose was further divided into G1 (no radiation), G2 (> 0 Mrad to < 5 Mrad), G3 (≥ 5 Mrad to < 10 Mrad), and G4 (≥ 10 Mrad). RESULTS: A total of 5,329 THAs were revised, 1,290 of which were due to aseptic loosening. Total radiation dose, face asymmetry, and stabilization treatments were found to significantly affect implant survival. G1 had the highest revision risk for any reason and for aseptic loosening and G3 and G4 the lowest. Compared with G1, the adjusted hazard ratio for G2 was 0.74 (95% confidence interval (CI) 0.64 to 0.86), G3 was 0.36 (95% CI 0.30 to 0.43), and G4 was 0.38 (95% CI 0.31 to 0.47). The cumulative incidence of revision for aseptic loosening at 12 years was 0.52 and 0.54 per 100 THAs for G3 and G4, respectively, compared with 1.95 per 100 THAs in G1. Asymmetrical liners had a lower revision risk due to aseptic loosening and reasons other than aseptic loosening compared with symmetric (flat) liners. In G3 and G4, stabilization with vitamin E and heating above melting point performed best. CONCLUSION: Polyethylene liners with a total radiation dose of ≥ 5 Mrad, an asymmetrical liner face, and stabilization with heating above the melting point demonstrate best survival. Cite this article: Bone Joint J 2020;102-B(1):90-101.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Davis ET, Pagkalos J, Kopjar B, et al. Polyethylene manufacturing characteristics have a major effect on the risk of revision surgery in cementless and hybrid total hip arthroplasties. Bone Joint J. 2020. PMID: 31888362. doi:10.1302/0301-620X.102B1.BJJ-2019-0779.R1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。