全髋关节置换术中下肢延长的临界限度:术中神经监测研究

Critical Limit of Lower-Extremity Lengthening in Total Hip Arthroplasty: An Intraoperative Neuromonitorization Study.
PubMed 文献 · J Bone Joint Surg Am · 2020-01-01

本研究在16例初次全髋关节置换术中采用术中神经监测,评估下肢延长对神经结构的影响及延长临界限度与人体测量学指标的关系。结果显示神经延长的临界限度与股骨长度、ASIS-MM距离和身高正相关,相对于股骨长度5%和ASIS-MM距离2.6%的延长被视为临界。

要点

  • 神经延长的安全限度平均为14.9±6.2 mm,临界限度平均为22.4±5.6 mm。
  • 临界延长与股骨长度(r=0.782)、ASIS-MM距离(r=0.811)和身高(r=0.835)正相关。
  • 临界限度取决于术中神经监测预设的阈值。

核心结论

  • 神经延长的临界限度与人体测量学指标直接相关,相对于股骨长度5%和ASIS-MM距离2.6%的延长为临界,但取决于IONM预设阈值。

研究设计与样本

  • 前瞻性术中神经监测研究,纳入16例接受初次全髋关节置换术的患者。

主要发现

  • 所有病例均在腓深神经达到初始IONM变化(安全限度)和最大允许IONM变化(临界限度)。
  • 安全限度平均为14.9±6.2 mm(相对于股骨长度3%,相对于ASIS-MM距离1.7%),临界限度平均为22.4±5.6 mm(相对于股骨长度5%,相对于ASIS-MM距离2.6%)。
  • 当腓深神经达到临界限度时,胫神经MEP波幅平均下降27%(95%CI 22.1%至32.7%),股神经MEP波幅平均下降12%(95%CI 6.9%至18.1%)。
  • 临界延长量与股骨长度(r=0.782;p<0.001)、ASIS-MM距离(r=0.811;p<0.001)和身高(r=0.835;p=0.001)呈正相关。
  • 临界延长量与胫神经和股神经的波幅下降之间无相关性。

关键数值与推荐

  • 术中神经监测临界限度定义为:SEP或MEP波幅较基线下降≥50%,或SEP潜伏期延长≥10%。

临床怎么用

  • 在全髋关节置换术中行下肢延长时,可参考相对于股骨长度5%和ASIS-MM距离2.6%的临界限度,并结合术中神经监测预设阈值进行个体化评估。

局限性

  • 研究样本量较小(16例),证据等级为预后Level IV。
  • 临界限度依赖于预先设定的IONM阈值。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The effect of limb lengthening on neural structures was assessed with use of intraoperative neuromonitoring (IONM) during primary total hip arthroplasty (THA). The relationship between the critical limit of lengthening and anthropometric measurements was evaluated. METHODS: Motor evoked potentials (MEPs) from the deep peroneal nerve (tibialis anterior muscle), tibial nerve (gastrocnemius muscle), and femoral nerve (quadriceps muscle), as well as somatosensory evoked potentials (SEPs) from the posterior tibial nerve, were recorded in 16 patients undergoing THA. Height, weight, the distance between the anterior superior iliac spine and the medial malleolus (ASIS-MM distance), and the total femoral length were measured preoperatively. Lower-extremity traction was performed after resection of the femoral head, and the amount of extremity lengthening was measured with use of an image intensifier. A maximum of 50% reduction in any one of the SEP or MEP amplitudes or a 10% increase in the SEP latency were considered to be indicative of the critical limit of lengthening. RESULTS: Initial IONM changes (indicating the safe limit of lengthening) and maximum allowed IONM changes (indicating the critical limit of lengthening) were reached in the deep peroneal nerve in all cases. The mean safe limit of lengthening (and standard deviation) was 14.9 ± 6.2 mm (3% relative to femoral length and 1.7% relative to ASIS-MM distance), whereas the critical limit of lengthening was exceeded at a mean of 22.4 ± 5.6 mm (5% relative to femoral length and 2.6% relative to ASIS-MM distance). When the critical limit was reached in the deep peroneal nerve, the mean decrease in MEP amplitudes was 27% (95% confidence interval [CI], 22.1% to 32.7%) for the tibial nerve and 12% (95% CI, 6.9% to 18.1%) for the femoral nerve. There was a positive correlation between critical lengthening and femoral length (r = 0.782; p < 0.001), ASIS-MM distance (r = 0.811; p < 0.001), and height (r = 0.835; p = 0.001). No correlation existed between the critical lengthening amount and the decrease in amplitude in the tibial and femoral nerves. CONCLUSIONS: The critical limit of nerve lengthening was directly correlated with anthropometric measurements. Nerve lengthening of 5% relative to femoral length and of 2.6% relative to ASIS-MM distance was found to be critical; however, these limits depended on the predetermined threshold values for IONM. LEVEL OF EVIDENCE: Prognostic Level IV. See Instructions for Authors for a complete description of levels of evidence.