双动髋臼假体3年迁移与磨损的放射立体分析

Migration and Wear of a Dual Mobility Acetabular Construct at 3 Years Measured by Radiostereometric Analysis.
PubMed 文献 · J Arthroplasty · 2019-11-14

本研究对27例非骨水泥型解剖双动髋臼假体(ADM)患者进行3年放射立体分析,评估髋臼杯迁移和聚乙烯磨损。结果显示2年时平均近端平移低于0.2 mm阈值,3年为0.16 mm(SD 0.31);入床期后年磨损率为0.02 mm/y,低于0.1 mm/y阈值;患者满意度96%,Oxford 12髋评分从术前21(SD 7)改善至术后3年43(SD 7)。结论提示ADM杯迁移稳定、磨损率低,双动和解剖设计未增加固定或磨损风险。

要点

  • ADM杯3年近端平移0.16 mm(SD 0.31),低于0.2 mm的长期生存可接受阈值。
  • 入床期后年磨损率0.02 mm/y,低于0.1 mm/y阈值,且杯迁移与聚乙烯磨损无关联。
  • 3年患者满意度96%,Oxford 12髋评分从21(SD 7)改善至43(SD 7)。

核心结论

  • ADM杯3年迁移稳定,提示无菌性松动风险低;入床期后年磨损率低,双动和解剖设计未增加固定或磨损风险。

研究设计与样本

  • 前瞻性队列研究,纳入27例接受非骨水泥型ADM髋臼假体植入的患者,在3年内进行6次放射立体分析(RSA)随访。

主要发现

  • 3年时平均近端平移为0.16 mm(SD 0.31),2年时已低于0.2 mm阈值。
  • 平均矢状面旋转为0.29度(SD 1.03),女性受试者更大(P < .001)。
  • 入床期后年磨损率为0.02 mm/y,低于0.1 mm/y阈值;杯迁移与聚乙烯磨损无关联。
  • 3年患者满意度为96%,Oxford 12髋评分从术前21(SD 7)改善至43(SD 7)。

关键数值与推荐

  • 近端平移阈值:2年时0.2 mm,与可接受的长期生存相关;本研究3年值为0.16 mm(SD 0.31)。
  • 年磨损率阈值:0.1 mm/y;本研究入床期后年磨损率为0.02 mm/y。

临床怎么用

  • ADM杯在3年内迁移稳定且磨损率低,支持其用于低脱位风险需求的髋臼重建。
  • 女性患者矢状面旋转更大(P < .001),术后随访中可关注杯旋转变化。

局限性

  • 样本量27例,结论需更大样本和更长随访验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The anatomic dual mobility (ADM) acetabular component was introduced because of previously described low dislocation rates for this type of construct. The shape of the anatomic cup and the motion of polyethylene liner may have implications for acetabular cup fixation and polyethylene liner wear; therefore, the purpose of this study was to assess the migration patterns and wear rates of the ADM component using radiostereometric analysis. METHODS: Uncemented ADM acetabular components were implanted in 27 patients. Radiostereometric analysis exams were taken at 6 follow-up visits over 3 years. Proximal translation and sagittal rotation of the cup and polyethylene total wear and wear rates were calculated. Oxford 12 Hip scores and satisfaction were recorded. RESULTS: Mean proximal translation was below the 0.2 mm threshold at 2 years associated with acceptable long-term survivorship (0.16 mm [standard deviation {SD} 0.31] at 3 years). Mean sagittal rotation was 0.29 degrees (SD 1.03) and was greater in female subjects (P < .001). Following bedding-in, the annual wear rate was 0.02 mm/y, below the 0.1 mm/y threshold. There was no association between cup migration and polyethylene wear. Patient satisfaction at 3 years was 96%. Mean Oxford 12 Hip scores improved from 21 (SD 7) preoperatively to 43 (SD 7) 3 years postoperatively. CONCLUSIONS: The ADM cup demonstrated stable migration at 3 years indicating low risk for aseptic loosening. Bedding-in in the first year was followed by low annual wear rates. These finding suggest no increased risk of the dual mobility and anatomic design on fixation or wear.