文献速递关节外科创伤骨科

模块化锥形开槽柄用于Paprosky III/IV或Vancouver B2/B3股骨翻修的结果

Radiographic and Clinical Outcomes of Modular Tapered Fluted Stems for Femoral Revision for Paprosky III and IV Femoral Defects or Vancouver B2 and B3 Femoral Fractures.
PubMed 文献 · J Arthroplasty · 2019-11-30

BACKGROUND: Extensive femoral bone loss poses a challenge in revision total hip arthroplasty (rTHA). Many techniques have been developed to address this problem including fully porous cylindrical stems, impaction bone grafting, and cementation of long stems, which have had varied…

要点

  • MTFS在Paprosky 3/4或Vancouver B2/B3股骨翻修中至少2年随访显示优良结果。
  • 中位下沉仅1.4 mm,95%股骨柄末次随访仍原位。
  • 以无菌性失败为终点,MTFS生存率达98.4%。

核心结论

  • 在Paprosky 3、4型股骨缺损或Vancouver B2、B3型骨折患者中,模块化锥形开槽柄行股骨翻修至少2年随访显示优良结果。

研究设计与样本

  • 回顾性查询临床数据库,纳入接受MTFS翻修且Paprosky 3/4股骨骨缺损或Vancouver B2/B3假体周围股骨骨折的患者;无2年随访者被邀请回院行X线评估并完成临床问卷。

主要发现

  • 中位下沉1.4 mm(范围0-21);全因翻修率16.3%(21例)。
  • 翻修原因:不稳定10例、感染6例、无菌性松动1例、假体周围股骨骨折1例、其他原因3例。
  • 7例(5.4%)行股骨柄翻修,最常见原因为感染(5例),另2例分别为Paprosky 4型股骨无菌性松动和假体周围股骨骨折。
  • 以无菌性失败为终点的锥形模块化开槽柄生存率为98.4%;末次随访平均HOOS-JR评分73,VR-12躯体与心理评分分别为32.8和52.2。

临床怎么用

  • 对于Paprosky 3/4型股骨缺损或Vancouver B2/B3型假体周围股骨骨折的翻修患者,MTFS可作为改善股骨侧固定的可选方案。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Extensive femoral bone loss poses a challenge in revision total hip arthroplasty (rTHA). Many techniques have been developed to address this problem including fully porous cylindrical stems, impaction bone grafting, and cementation of long stems, which have had varied success. Modular tapered fluted femoral stems (MTFS) show favorable results. We sought to determine the minimum 2-year radiographic and clinical performance of MTFS in rTHA in a population with extensive proximal femoral bone loss. METHODS: Our clinical database was queried retrospectively for all patients who underwent rTHA with an MTFS. We included patients with Paprosky 3 and 4 femoral bone loss and patients with Vancouver B2 and B3 periprosthetic femur fractures. Patients without 2-year follow-up were invited to return to clinic for X-ray evaluation and to complete clinical questionnaires. We assessed distance of stem subsidence and presence of stem fixation on final X-ray. We recorded all-cause revision and survival of the stem at final follow-up. RESULTS: One hundred twenty-nine patients were available for follow-up. Average follow-up time was 3.75 years. One hundred twenty-two stems (95%) remained in place at final follow-up. Median subsidence was 1.4 mm (range 0-21). All-cause revision rate was 16.3% (21 patients). Of the hips revised, 10 were for instability, 6 for infection, 1 for aseptic loosening, and 1 for periprosthetic femur fracture. Three were revised for other reasons. The stem was revised in 7 patients (5.4%), and the most common reason for stem revision was infection (5 patients). The other 2 stems were revised for aseptic loosening in a Paprosky 4 femur and periprosthetic femur fracture. Survival of tapered modular fluted stems with aseptic failure as an endpoint was 98.4%. The mean Hip disability and Osteoarthritis Outcome Score, Joint Replacement score at final follow-up was 73, and mean Veterans Rand 12 item health survey physical and mental scores were 32.8 and 52.2, respectively. CONCLUSION: In patients with Paprosky 3, 4 femoral defects or Vancouver type B2, B3 fractures, modular tapered fluted stems for femoral revision show excellent outcomes at minimum 2-year follow-up.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Otero JE, Martin JR, Rowe TM, et al. Radiographic and Clinical Outcomes of Modular Tapered Fluted Stems for Femoral Revision for Paprosky III and IV Femoral Defects or Vancouver B2 and B3 Femoral Fractures. J Arthroplasty. 2019. PMID: 31870582. doi:10.1016/j.arth.2019.11.039

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。