金属对聚乙烯与陶瓷对聚乙烯全髋关节置换模拟器中的钴铬离子释放及其细胞与微生物相关性
该模拟器研究比较了金属对聚乙烯(MoP)与陶瓷对聚乙烯(CoP)全髋关节置换构型在100万次循环后钴(Co)和铬(Cr)离子释放水平,并评估这些离子对人脂肪间充质干细胞(AMSC)的细胞毒性及对表皮葡萄球菌生长的影响。结果显示MoP构型释放大量CoCr碎屑,Co离子对AMSC呈剂量依赖性毒性,但模拟器观察到的Co浓度未显著影响表皮葡萄球菌生长。
要点
- MoP构型在理想模拟条件下释放大量CoCr碎屑,提示基线风险。
- Co离子对人AMSC呈剂量依赖性毒性,碎屑水平可具细胞毒性。
- 模拟器Co浓度未增强或抑制表皮葡萄球菌生长。
核心结论
- MoP构型在理想条件下产生大量CoCr碎屑,存在基线风险,可能被宿主因素或手术技术不完美加剧。
- 模拟器条件产生的碎屑水平对AMSC具有细胞毒性。
- 这些浓度未增强或抑制表皮葡萄球菌生长,提示假体周围感染率升高与不良局部组织反应相关,可能与其他因素(如组织坏死)有关。
研究设计与样本
- 体外研究:10个髋关节模拟器构型(5个MoP和5个CoP)在牛血清中运行1,000,000次循环,评估CoCr浓度。
- 体外培养人脂肪间充质干细胞(AMSC)和表皮葡萄球菌,比较CoCr与惰性二氧化硅的细胞毒性和生长影响。
主要发现
- 1,000,000次循环后,MoP和CoP的平均Co浓度分别为2264和0.6 ng/mL(P < .001)。
- MoP和CoP的平均Cr浓度分别为217和4.3 ng/mL(P < .001)。
- MoP的平均Co:Cr比为10:1。
- Co离子对AMSC的毒性显著高于对照二氧化硅,呈剂量依赖性(P < .001)。
- 模拟器观察到的Co浓度未显著影响表皮葡萄球菌生长。
临床怎么用
- MoP全髋关节置换在理想模拟条件下即可释放大量CoCr碎屑,临床应关注基线风险。
- 模拟器条件产生的CoCr碎屑水平对AMSC具有细胞毒性,提示可能影响局部组织。
- 模拟器Co浓度未影响表皮葡萄球菌生长,假体周围感染率升高可能与其他因素(如组织坏死)相关。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The aims of this study were to determine the levels of cobalt (Co) and chromium (Cr) ions generated in simulators from metal-on-polyethylene (MoP) and ceramic-on-polyethylene (CoP) constructs. Furthermore, we aimed to investigate the cytotoxic effect of these ion levels on native tissues and their potential to modify periprosthetic joint infection risk. METHODS: We used in vitro culture of human adipose-derived mesenchymal stem cells (AMSCs) and Staphylococcus epidermidis cultures, respectively. Ten hip simulator constructs (5 MoP and 5 CoP) were assembled and run for 1,000,000 cycles in bovine serum and evaluated for CoCr concentration. Cytotoxicity and growth impact on AMSCs and S. epidermidis was compared between CoCr and inert silicon dioxide. RESULTS: After 1,000,000 cycles, mean MoP and CoP Co concentration was 2264 and 0.6 ng/mL, respectively (P < .001). Mean MoP and CoP Cr concentration was 217 and 4.3 ng/mL, respectively (P < .001). Mean MoP Co:Cr ratio was 10:1. Co ions were significantly more toxic to human AMSCs than control silicon dioxide in a dose-response manner (P < .001). S. epidermidis growth was not significantly impacted by Co concentrations observed in the simulators. CONCLUSION: MoP constructs built in ideal conditions generated substantial CoCr debris, highlighting a baseline risk with these implants that may be exacerbated by host factors or imperfect surgical technique. Evaluation of impact on AMSCs suggests that debris levels produced under simulator conditions can be cytotoxic. In addition, these concentrations did not potentiate or inhibit S. epidermidis growth, suggesting that elevated periprosthetic joint infection rates with adverse local tissue reaction are related to other factors potentially associated with tissue necrosis.
引用
- Wyles CC, Paradise CR, Masters TL, et al. Cobalt and Chromium Ion Release in Metal-on-Polyethylene and Ceramic-on-Polyethylene THA: A Simulator Study With Cellular and Microbiological Correlations. J Arthroplasty. 2019. PMID: 31852609. doi:10.1016/j.arth.2019.11.031
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。