文献速递关节外科

骨质量对内侧骨水泥型单髁膝关节置换术后胫骨假体迁移的影响

The Effect of Bone Quality on Tibial Component Migration in Medial Cemented Unicompartmental Knee Arthroplasty. A Prospective Cohort Study Using Dual X-Ray Absorptiometry and Radiostereometric Analysis.
PubMed 文献 · J Arthroplasty · 2019-10-24

结果显示两组假体迁移均较低,且不受术前全身骨密度或术后假体周围骨密度变化影响。

要点

  • 正常骨密度组所有感兴趣区骨密度较低于低骨密度组11%至15%。
  • 两组术后假体周围骨密度在ROI 1-3均随时间下降,术膝与对侧膝下降相似。
  • 12至24个月时两组最大总点迁移分别为0.03 mm和0.02 mm,迁移未受骨密度变化影响。

核心结论

  • 骨水泥型内侧胫骨UKA假体迁移至24个月均较低,且不受术前全身骨密度或术后假体周围骨密度变化影响,提示尽管近端胫骨骨密度存在指数差异,长期固定仍良好。

主要发现

  • 正常全身骨密度组所有感兴趣区骨密度在整个随访期间较低全身骨密度组高11%至15%。
  • 两组假体周围骨密度在ROI 1-3随时间下降,术膝与对侧膝术前至24个月骨密度下降相似。
  • 12至24个月,正常骨密度组最大总点迁移0.03 mm(95%CI -0.01, 0.08),低骨密度组0.02 mm(95%CI -0.03, 0.07)。
  • 迁移随时间的变化不受假体周围骨密度变化影响。

临床怎么用

  • 对于内侧骨水泥型UKA,术前全身骨密度及术后假体周围骨密度变化可能不显著影响胫骨假体中期迁移,但需结合个体情况综合评估。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Periprosthetic bone mineral density (BMD) may influence implant fixation and subsequent loosening. Unicompartmental knee arthroplasty (UKA) restores normal knee kinematics and load distribution to the surrounding bone. We studied the influence of systemic and periprosthetic BMD of the proximal tibia on migration of the tibial component of cemented medial UKA. METHODS: The cohort was dichotomized into a normal BMD group (T-score ≥ -1; n = 37) and a low BMD group (T-score < -1; n = 28) according to World Health Organization criteria. BMD of the proximal tibia and migration of the tibial component were measured with dual X-ray absorptiometry scans and stereoradiographs with 2-year follow-up. RESULTS: Patients with normal systemic BMD had an 11% to15% higher BMD in all regions of interest (ROIs) compared to patients with low systemic BMD throughout follow-up. Over time, a decrease in periprosthetic BMD in ROI 1-3 was seen for both groups. The operated knees and contralateral knees showed a similar reduction in BMD in all ROIs between preoperative and 24 months. Between 12 and 24 months, the normal BMD group migrated (maximal total point motion) 0.03 mm (95% confidence interval, -0.01, 0.08) and the low BMD group migrated 0.02 mm (95% confidence interval, -0.03, 0.07). Migration over time was not influenced by change in periprosthetic BMD. CONCLUSION: Migration of cemented medial tibial UKA was low until 24 months and was neither affected by preoperative systemic BMD nor affected by postoperative change in periprosthetic BMD. This suggests good long-term fixation despite an index difference in proximal tibial BMD.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Koppens D, Rytter S, Dalsgaard J, et al. The Effect of Bone Quality on Tibial Component Migration in Medial Cemented Unicompartmental Knee Arthroplasty. A Prospective Cohort Study Using Dual X-Ray Absorptiometry and Radiostereometric Analysis. J Arthroplasty. 2019. PMID: 31732369. doi:10.1016/j.arth.2019.10.027

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。