术中数字放射摄影确定并实现全髋关节置换术的目标肢体长度和偏移

Determining and Achieving Target Limb Length and Offset in Total Hip Arthroplasty Using Intraoperative Digital Radiography.
PubMed 文献 · J Arthroplasty · 2019-10-07

结果显示,100%患者肢体长度在目标5 mm内,97.1%患者偏移在目标5 mm内,术中与术后测量无显著差异,平均随访4.2年无脱位或再手术。

要点

  • 目标肢体长度和偏移不一定等于对侧,需个体化确定。
  • 术中数字放射摄影可帮助术者将肢体长度和偏移控制在目标5 mm内。
  • 单侧髋骨关节炎患者中,该方法临床结果优良,无脱位或再手术。

核心结论

  • 大多数全髋关节置换术患者可确定目标肢体长度和偏移,且目标不总等于对侧。
  • 术中数字放射摄影能使术者将目标肢体长度和偏移精确控制在5 mm内,临床结果优良。

研究设计与样本

  • 术前、术中、术后测量肢体长度和偏移,术后平均随访4.2年。

主要发现

  • 肢体长度在目标5 mm内的患者占100%,偏移在目标5 mm内的患者占97.1%。
  • 目标测量值与对侧相差>5 mm的比例:肢体长度2.0%,偏移2.9%。
  • 术中与术后肢体长度测量无显著差异(P = .261),偏移测量无显著差异(P = .747)。
  • 末次随访无脱位或再手术,平均髋关节功能障碍和骨关节炎结局评分关节置换版为95.78。

关键数值与推荐

  • 偏移定义为平行于泪滴间线、从坐骨边缘(约在闭孔下缘水平)至股骨皮质边缘(通常在颈切除水平或略下方)的线段长度。
  • 目标肢体长度根据患者感知和疾病严重程度确定,目标偏移等于对侧肢体。

临床怎么用

  • 术前根据患者感知和疾病严重程度个体化确定目标肢体长度,目标偏移通常参考对侧。
  • 术中可使用数字放射摄影测量并调整,直至达到目标且髋关节稳定。

局限性

  • 研究为单中心、同质队列(单侧髋骨关节炎),结论外推需谨慎。
  • 未报告术中调整的具体次数或相关并发症。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Achieving appropriate limb length and offset in total hip arthroplasty (THA) is challenging. Target limb length and offset may not always mean equal radiographic measurements bilaterally. The goal of this study is to introduce a method for determining as well as achieving target limb length and offset using digital radiographic measurements. METHODS: One hundred and two consecutive patients with unilateral hip osteoarthritis undergoing primary THA in the lateral decubitus position were included. Limb length and offset were measured on anterior-posterior pelvic radiographs preoperatively, intraoperatively, and postoperatively. Offset was defined as the length of a line parallel to the inter-teardrop line, extending from the edge of the ischium, at about the lower border of the ipsilateral obturator foramen, to the edge of the femoral cortex, usually at, or just below, the neck resection level. Target limb length was determined for each patient based on patient perception and severity of disease. Target offset equaled the contralateral limb. Using intraoperative digital radiography, adjustments were made until targets were achieved and the hip was stable. Patients were followed for an average of 4.2 years postoperatively. RESULTS: Limb length was within 5 mm of target measurements in 100% of patients and offset was within 5 mm of targets in 97.1%. Target measurements differed by >5 mm from the contralateral side in 2.0% of limb length and 2.9% of offset measurements. There were no significant differences between intraoperative and postoperative limb length (P = .261) or offset (P = .747) measurements. At final follow-up, there were no dislocations or reoperations and average Hip disability and Osteoarthritis Outcome Score for Joint Replacement was 95.78. CONCLUSION: Target limb length and offset goals can be determined for most patients undergoing THA. Targets are not always equal to the contralateral side. Intraoperative digital radiography can allow surgeons to accurately achieve target limb length and offset to within 5 mm in a homogenous cohort of patients with unilateral hip osteoarthritis with excellent clinical outcomes.