文献速递关节外科

全膝关节置换术后金属沉积与假体周围组织金属浓度的尸检取材分析

Postmortem Retrieval Analysis of Metallosis and Periprosthetic Tissue Metal Concentrations in Total Knee Arthroplasty.
PubMed 文献 · J Arthroplasty · 2019-08-19

本研究在尸检时采集20例全膝关节置换(TKA)假体及5个不同部位的组织样本,用ICP-MS测定钴(Co)、铬(Cr)、钛(Ti)浓度,并行滑液细胞因子分析和假体损伤评估。结果显示组织中Co、Cr、Ti中位浓度分别为16 ppb、46 ppb、9.8 ppb,采集部位与金属浓度无关,20例中12例存在金属沉积但金属浓度与金属沉积无相关性。

要点

  • 尸检取材的TKA组织中Co、Cr、Ti中位浓度分别为16 ppb、46 ppb、9.8 ppb。
  • 组织采集部位与Co、Cr、Ti浓度均无关联(P=.979、.712、.854)。
  • 20例中12例存在金属沉积,但金属浓度与金属沉积无相关性。

核心结论

  • 尸检取材的TKA组织中可观察到Co、Cr、Ti浓度升高。
  • 研究结果不支持组织金属浓度与炎症细胞因子相关的假设。

研究设计与样本

  • 在尸检时采集20例TKA系统及来自5个不同部位的组织样本。
  • 采用电感耦合等离子体质谱(ICP-MS)测定Co、Cr、Ti浓度。
  • 采用磁性Luminex筛选试验进行滑液细胞因子分析,评估股骨部件损伤并对组织金属沉积进行视觉评分。

主要发现

  • 组织中Co、Cr、Ti的中位浓度分别为16 ppb、46 ppb、9.8 ppb。
  • 组织采集部位与Co(P=.979)、Cr(P=.712)、Ti(P=.854)浓度均无关联。
  • 20例尸检取回的TKA中12例存在金属沉积,但Co(P=.48)、Cr(P=.89)、Ti(P=.60)浓度与金属沉积无相关性。
  • Co浓度升高与肿瘤坏死因子α降低相关(ρ=-0.56,P=.01),与白细胞介素1β降低相关(ρ=-0.48,P=.03)。
  • Cr浓度升高与肿瘤坏死因子α降低(ρ=-0.47,P=.03)、白细胞介素6降低(ρ=-0.43,P=.04)、巨噬细胞炎症蛋白3α降低(ρ=-0.47,P=.03)相关。

临床怎么用

  • 组织金属浓度分析可选择任意采集部位,因不同部位间浓度无差异。
  • 金属沉积的存在不能由组织金属浓度推断,反之亦然。
  • 金属浓度与炎症细胞因子的关联方向与预期相反,解读时需谨慎。

局限性

  • 本研究为尸检取材的回顾性分析,样本量为20例。
  • 研究结果需未来前瞻性研究进一步验证。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The purpose of this study is to determine the preferred sampling location for tissue analysis in total knee arthroplasty (TKA) and to evaluate metal concentrations, inflammatory cytokines, component damage, and tissue metallosis. METHODS: Twenty TKA systems were collected at necropsy along with tissue samples from 5 distinct locations. Inductively coupled plasma mass spectrometry (ICP-MS) analysis was performed to determine cobalt (Co), chromium (Cr), and titanium (Ti) concentrations. Synovial fluid cytokine analysis was preformed using a Magnetic Luminex Screening Assay. Femoral components were assesed for damage and tissues were visually scored for metallosis. RESULTS: The median metal concentrations were 16 ppb for Co, 46 ppb for Cr, and 9.8 ppb for Ti. There was no association between the tissue collection site and the metal concentration for Co (P = .979), Cr (P = .712), or Ti (P = .854). Twelve of 20 of the necropsy-retrieved TKAs had metallosis, but there was no correlation between Co (P = .48), Cr (P = .89), or Ti (P = .60) concentration and metallosis. Increased Co was associated with decreased tumor necrosis factor alpha (ρ = -0.56, P = .01) and interleukin 1 beta (ρ = -0.48, P = .03). Increased Cr was associated with decreased tumor necrosis factor alpha (ρ= -0.47, P = .03), interleukin 6 (ρ= -0.43, P = .04), and macrophage inflammatory protein 3 alpha (ρ= -0.47, P = .03). CONCLUSION: We observed elevated Co, Cr, and Ti concentrations in tissue from necropsy-retrieved TKA. Our findings did not support the hypothesis that tissue metal concentrations were associated with inflammatory cytokines. The results of this research will be useful for the design of future prospective studies.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Arnholt CM, White JB, Lowell JA, et al. Postmortem Retrieval Analysis of Metallosis and Periprosthetic Tissue Metal Concentrations in Total Knee Arthroplasty. J Arthroplasty. 2019. PMID: 31699531. doi:10.1016/j.arth.2019.08.038

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。