文献速递关节外科

基于模型放射立体测量分析评估现代全膝关节置换术的体内磨损

In vivo wear measurement in a modern total knee arthroplasty with model-based radiostereometric analysis.
PubMed 文献 · Bone Joint J · 2019-01-01

这项回顾性研究采用基于模型的放射立体测量分析(MBRSA)测量了101例全膝关节置换术(TKA)患者聚乙烯(PE)内衬的短期体内线性和体积磨损。术后及6、12、24个月进行仰卧非负重RSA检查,部分患者在12和24个月进行站立负重RSA检查。线性磨损率为仰卧0.015 mm/年、站立0.220 mm/年;体积磨损率为仰卧10.3 mm³/年、站立39.3 mm³/年,磨损主要发生在关节内侧,负重成像大幅提高测量可靠性。

要点

  • MBRSA可在短期随访中评估TKA的体内体积磨损。
  • 站立负重RSA测得的磨损率高于仰卧非负重测量。
  • 磨损主要发生于关节内侧,患者 demographics 与功能评分与磨损无显著关联。

核心结论

  • MBRSA可用于全膝关节置换术聚乙烯内衬的短期体内体积磨损评估。

研究设计与样本

  • 回顾性研究,纳入101例全膝关节置换术患者,术后及6、12、24个月行仰卧非负重RSA,部分患者12和24个月行站立负重RSA。

主要发现

  • 线性磨损率为仰卧0.015 mm/年、站立0.220 mm/年。
  • 体积磨损率为仰卧10.3 mm³/年、站立39.3 mm³/年。
  • 磨损主要发生在关节内侧,负重成像大幅提高测量可靠性。
  • 体积磨损的临床精度为34 mm³。
  • 患者年龄、性别、体重指数和牛津膝关节评分与磨损无显著关联。

关键数值与推荐

  • 体积磨损临床精度为34 mm³。

临床怎么用

  • 对于全膝关节置换术后患者,可采用MBRSA进行短期体内磨损监测,负重成像可提高测量可靠性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: A retrospective study was conducted to measure short-term in vivo linear and volumetric wear of polyethylene (PE) inserts in 101 total knee arthroplasty (TKA) patients using model-based radiostereometric analysis (MBRSA). PATIENTS AND METHODS: Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear. RESULTS: Linear wear occurred at 0.015 mm/year (supine) and 0.220 mm/year (standing). Volumetric wear occurred at 10.3 mm3/year (supine) and 39.3 mm3/year (standing). Wear occurred primarily on the medial side of the joint. Weightbearing imaging greatly improved the reliability of measurement. Clinical precision of volumetric wear was 34 mm3. No significant associations were found between patient demographics or function scores and measured wear. CONCLUSION: In vivo volumetric wear of TKAs can be assessed at short-term follow-up using MBRSA. Cite this article: Bone Joint J 2019;101-B:1348-1355.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Gascoyne T, Parashin S, Teeter M, et al. In vivo wear measurement in a modern total knee arthroplasty with model-based radiostereometric analysis. Bone Joint J. 2019. PMID: 31674253. doi:10.1302/0301-620X.101B11.BJJ-2018-1447.R2

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。