cemented维生素E稳定高交联聚乙烯组件的随机双盲非劣效性试验:反向混合全髋关节置换术中的迁移评估
这项双盲随机对照非劣效性试验纳入42例骨关节炎患者,比较骨水泥固定维生素E掺杂高交联聚乙烯(VEPE)组件与常规聚乙烯组件在反向混合全髋关节置换术中的安全性。主要终点为术后两年RSA测量的组件近端迁移,VEPE组近端迁移显著高于对照组,平均差0.21 mm(95%CI 0.05至0.37;p=0.013),未达到非劣效性;股骨头穿透更低,临床结果无差异。
要点
- VEPE组两年近端迁移较对照组平均高0.21 mm(95%CI 0.05至0.37;p=0.013),未达非劣效性。
- VEPE组总迁移也显著更高,但股骨头穿透更低。
- 两组患者报告结局测量无差异。
核心结论
- 两年时,骨水泥VEPE组件虽股骨头穿透低且临床结果优异,但近端迁移超过RSA提出的安全阈值,未能达到相对于常规植入物的非劣效性。
- VEPE组件的早期近端迁移模式是持续监测的理由,但近端迁移的具体阈值及后期失败风险无法确定,需进一步研究。
研究设计与样本
- 双盲、非劣效性随机对照试验,纳入42例骨关节炎患者(21男,21女),平均年龄67岁(sd 5)。
- 受试者按1:1随机接受反向混合全髋关节置换术,使用几何形状相同的骨水泥固定氩气伽马灭菌聚乙烯组件(对照组)或VEPE组件。
- 主要终点为术后两年RSA测量的组件近端迁移;次要终点包括组件总迁移、股骨头穿透及患者报告结局测量。
主要发现
- 主要终点分析纳入对照组19个植入物和VEPE组18个植入物。
- VEPE组组件近端迁移持续存在,两年时平均差0.21 mm(95%CI 0.05至0.37;p=0.013),VEPE组显著高于对照组。
- VEPE组总迁移也显著高于对照组,但股骨头穿透低于对照组。
- 两组临床结果无差异。
关键数值与推荐
- 主要终点:术后两年RSA测量的组件近端迁移,非劣效性比较。
- VEPE组与对照组两年近端迁移平均差0.21 mm(95%CI 0.05至0.37;p=0.013),VEPE组更高。
临床怎么用
- VEPE组件股骨头穿透低且临床结果优异,但早期近端迁移高于常规聚乙烯组件,未达非劣效性。
- VEPE组件的早期近端迁移模式是持续监测的理由。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: Radiostereometric analysis (RSA) studies of vitamin E-doped, highly crosslinked polyethylene (VEPE) liners show low head penetration rates in cementless acetabular components. There is, however, currently no data on cemented VEPE acetabular components in total hip arthroplasty (THA). The aim of this study was to evaluate the safety of a new cemented VEPE component, compared with a conventional polyethylene (PE) component regarding migration, head penetration, and clinical results. PATIENTS AND METHODS: We enrolled 42 patients (21 male, 21 female) with osteoarthritis and a mean age of 67 years (sd 5), in a double-blinded, noninferiority, randomized controlled trial. The subjects were randomized in a 1:1 ratio to receive a reverse hybrid THA with a cemented component of either argon-gas gamma-sterilized PE component (controls) or VEPE, with identical geometry. The primary endpoint was proximal implant migration of the component at two years postoperatively measured with RSA. Secondary endpoints included total migration of the component, penetration of the femoral head into the component, and patient-reported outcome measurements. RESULTS: In total, 19 control implants and 18 implants in the VEPE group were analyzed for the primary endpoint. We found a continuous proximal migration of the component in the VEPE group that was significantly higher with a difference at two years of a mean 0.21 mm (95% confidence interval (CI) 0.05 to 0.37; p = 0.013). The total migration was also significantly higher in the VEPE group, but femoral head penetration was lower. We found no difference in clinical outcomes between the groups. CONCLUSION: At two years, this cemented VEPE component, although having a low head penetration and excellent clinical results, failed to meet noninferiority compared with the conventional implant by a proximal migration above the proposed safety threshold of RSA. The early proximal migration pattern of the VEPE component is a reason for continued monitoring, although a specific threshold for proximal migration and risk for later failure cannot be defined and needs further study. Cite this article: Bone Joint J 2019;101-B:1192-1198.
引用
- Sköldenberg OG, Rysinska AD, Chammout G, et al. A randomized double-blind noninferiority trial, evaluating migration of a cemented vitamin E-stabilized highly crosslinked component compared with a standard polyethylene component in reverse hybrid total hip arthroplasty. Bone Joint J. 2019. PMID: 31564144. doi:10.1302/0301-620X.101B10.BJJ-2019-0456.R2