文献速递关节外科

混合与同厂双动关节内衬的假体头脱位所需杠杆力无差异

No Difference in Force Required for Intraprosthetic Dislocation of Mixed Manufacturer vs Same Manufacturer Dual Mobility Articulations.
PubMed 文献 · J Arthroplasty · 2019-09-30

BACKGROUND: To avoid the morbidity of removing well-fixed implants during revision surgery, the off-label practice of mixing femoral heads with dual mobility (DM) polyethylene liners from different manufacturers is commonly performed.…

要点

  • 裙边存在与最大LOF增加63.8 N相关,头材料(陶瓷vs金属)不影响LOF。

核心结论

  • 同厂与混合厂商内头从双动关节内衬中杠杆脱出所需力无差异,提示在固定良好的股骨柄上混合厂商放置双动关节不应增加假体内不稳风险。

研究设计与样本

  • 2010年1月至2018年7月回收168个双动关节内衬,排除灾难性磨损(n=14)、既往杠杆脱出(n=17)和法律诉讼病例(n=8),最终129个标本进行单轴加载至杠杆脱出失败。
  • 比较同厂与混合厂商回收标本的最大杠杆脱出力(LOF),使用Student t检验,并用多变量回归分析植入时长、头尺寸、头材料、裙边存在对LOF的影响。

主要发现

  • 裙边存在与最大LOF增加63.8 N相关。
  • 头材料(陶瓷vs金属)不影响LOF。

关键数值与推荐

  • 裙边存在与最大LOF增加63.8 N相关。

临床怎么用

  • 在固定良好的股骨柄上放置双动关节时,混合厂商内头与内衬不应增加假体内不稳风险。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: To avoid the morbidity of removing well-fixed implants during revision surgery, the off-label practice of mixing femoral heads with dual mobility (DM) polyethylene liners from different manufacturers is commonly performed. The resistance to intraprosthetic dislocation, when the inner prosthetic head disengages from the polyethylene bearing, between mixed and same manufacturer constructs remains unknown. METHODS: Between January 2010 and July 2018, 168 DM liners were retrieved. Specimens were excluded for catastrophic wear (n = 14), previously levered-out (n = 17), and cases in legal proceedings (n = 8). Using a validated setup, 129 specimens were uniaxially loaded 100 mm from the femoral head until lever-out failure of the head from the liner. The difference in maximum lever-out force (LOF) was compared for same and mixed manufacturer retrievals (Student t-test). Multivariable regression analysis evaluated the influence of potential confounders (length of implantation, head size, head material, presence of skirt) on LOF. RESULTS: Ninety-seven same and 32 mixed manufacturer DM constructs were tested. The average LOF for same (272.6 ± 68.7 N) and mixed (299.2 ± 89.0 N) manufacturer specimens was not significantly different (P = .08). An inner head size of 22.2 mm was associated with 184.4-N increase in LOF (P < .001), the presence of a skirt was associated with 63.8-N increase in maximum LOF, and head material (ceramic vs metal) did not influence LOF. CONCLUSION: We found no difference in the force required to lever-out same and mixed manufacturer inner heads from DM liners, suggesting that mixing manufacturers when placing DM articulations on well-fixed femoral stems should not increase the risk of intraprosthetic instability.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Buller LT, Torres L, Baral EC, et al. No Difference in Force Required for Intraprosthetic Dislocation of Mixed Manufacturer vs Same Manufacturer Dual Mobility Articulations. J Arthroplasty. 2019. PMID: 31653465. doi:10.1016/j.arth.2019.09.039

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。