该体外研究用聚氨酯泡沫模型评估钴铬钼与氮化锆多层涂层胫骨假体在五种骨水泥下的固定强度,并分析骨水泥放置时机的影响。结果显示两种假体材料间无显著差异,早期至中期放置骨水泥对界面固定有益。

要点

  • 钴铬钼与氮化锆涂层胫骨托在五种骨水泥下固定强度无显著差异。
  • 骨水泥放置与假体植入在特定处理窗口早期至中期可能有益于界面固定。
  • 研究为体外标准化模型,共20组(每组n=5)及12组(每组n=5)时机测试。

核心结论

  • 钴铬钼与氮化锆多层涂层胫骨假体在五种常用骨水泥下植入-骨水泥-骨固定强度无显著差异。
  • 骨水泥放置和胫骨托植入在骨水泥特定处理窗口的早期至中期对全膝关节置换术界面固定有益。

研究设计与样本

  • 体外研究,使用合成聚氨酯泡沫模型,通过推出试验研究胫骨假体骨水泥固定后的固定强度。
  • 共20组(每组n=5):Vega PS钴铬钼胫骨与Vega PS氮化锆胫骨分别配合BonOs R、SmartSet HV、Cobalt HV、Palacos R和Surgical Simplex P五种骨水泥,采用中期骨水泥放置。
  • 共12组(每组n=5):使用BonOs R和SmartSet HV两种骨水泥,测试早期、中期和晚期三种骨水泥放置时间。

主要发现

  • 钴铬钼与氮化锆多层涂层Vega胫骨托在五种常用骨水泥下植入-骨水泥-骨固定强度无显著差异。
  • 骨水泥放置和胫骨托植入在骨水泥特定处理窗口的早期至中期对界面固定有益。

临床怎么用

  • 骨水泥放置和胫骨假体植入在骨水泥特定处理窗口的早期至中期可能有益于全膝关节置换术界面固定。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The objectives of this study are to (1) evaluate if there is a potential difference in cemented implant fixation strength between tibial components made out of cobalt-chromium (CrCoMo) and of a ceramic zirconium nitride (ZrN) multilayer coating and to (2) test their behavior with 5 different bone cements in a standardized in vitro model for testing of the implant-cement-bone interface conditions. We also analyzed (3) whether initial fixation strength is a function of timing of the cement apposition and component implantation by an early, mid-term, and late usage within the cement-specific processing window. METHODS: An in vitro study using a synthetic polyurethane foam model was performed to investigate the implant fixation strength after cementation of tibial components by a push-out test. A total of 20 groups (n = 5 each) was used: Vega PS CrCoMo tibia and Vega PS ZrN tibia with the bone cements BonOs R, SmartSet HV, Cobalt HV, Palacos R, and Surgical Simplex P, respectively, using mid-term cement apposition. Three different cement apposition times-early, mid-term, and late usage-were tested with a total of 12 groups (n = 5 each) with the bone cements BonOs R and SmartSet HV. RESULTS: There was no significant difference in implant-cement-bone fixation strength between CrCoMo and ZrN multilayer-coated Vega tibial trays tested with 5 different commonly used bone cements. CONCLUSION: Apposition of bone cements and tibial tray implantation in the early to mid of the cement-specific processing window is beneficial in regard to interface fixation in TKA.