文献速递脊柱外科

无缺口预弯棒用于脊柱畸形手术的力学分析

Mechanical Analysis of Notch-Free Pre-Bent Rods for Spinal Deformity Surgery.
PubMed 文献 · Spine (Phila Pa 1976) · 2020-01-01

本研究按ASTM F2193方法对钛合金(ϕ6.0 mm)和钴铬合金(ϕ5.5 mm)脊柱棒进行静态与动态四点弯曲测试,比较无缺口弯棒与传统缺口弯棒的力学性能。结果显示无缺口钴铬合金棒的极限载荷显著高于传统缺口钴铬合金棒和钛合金棒,动态测试中无缺口钴铬合金棒在2,500,000次循环最小力下的力/位移大于缺口钴铬合金棒。结论认为无缺口弯钴铬合金棒可能在脊柱畸形术后维持曲度并降低断裂风险。

要点

  • 无缺口钴铬合金棒极限载荷显著高于传统缺口钴铬合金棒和钛合金棒。
  • 静态四点弯曲中钴铬合金棒刚度显著高于钛合金棒。
  • 动态测试中无缺口钴铬合金棒在2,500,000次循环最小力下力/位移大于缺口钴铬合金棒。

核心结论

  • 无缺口弯钴铬合金棒可能在脊柱畸形术后维持曲度并降低断裂风险,或可克服传统缺口棒断裂和回弹问题。

研究设计与样本

  • 按ASTM F2193方法对商用钛合金(ϕ6.0 mm)和钴铬合金(ϕ5.5 mm)脊柱棒进行静态和动态四点弯曲实验研究。

主要发现

  • 静态四点弯曲测试中,钴铬合金弯棒刚度显著高于钛合金棒。
  • 无缺口钴铬合金棒极限载荷显著高于传统缺口钴铬合金棒和钛合金棒。
  • 动态四点弯曲测试中,无缺口钴铬合金棒在2,500,000次循环最小力下的力/位移大于缺口钴铬合金棒。

关键数值与推荐

  • 静态四点弯曲:钴铬合金弯棒刚度显著高于钛合金棒。
  • 无缺口钴铬合金棒极限载荷显著高于传统缺口钴铬合金棒和钛合金棒。
  • 动态四点弯曲:2,500,000次循环最小力下,无缺口钴铬合金棒力/位移大于缺口钴铬合金棒。

临床怎么用

  • 无缺口弯钴铬合金棒可能在脊柱畸形手术中维持曲度并降低断裂风险。

局限性

  • 原文未述局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

STUDY DESIGN: Experimental study of spinal rod as per the American Society for Testing Materials (ASTM) F2193 methodology for static and dynamic four-point bending. OBJECTIVE: The hypotheses underlying this study were that the notch-free, curved rod would have a significantly higher ultimate load and fatigue strength compared with conventional notched curved rods. This study aimed to analyze the mechanical properties of notch-free curved rods compared with conventional notched rods. SUMMARY OF BACKGROUND DATA: The goal of instrumented spinal fusion in the management of spinal deformities is to realign the spine and maintain the correction and stability in order to obtain arthrodesis. Although rod curvature could play an important role, intraoperative contouring of the straight rod induces notches into the rod, leading to decreased fatigue strength. METHODS: Commercially produced titanium alloy (ϕ6.0 mm) and cobalt chromium alloy (ϕ5.5 mm) spinal rods were assessed by four-point bending tests in accordance with the ASTM F2193. RESULTS: Static four-point bending tests for the curved spinal rods showed that cobalt chromium alloy rods had significantly higher stiffness compared with titanium alloy rods. Notch-free cobalt chromium alloy rods had a significantly higher ultimate load than the conventional notched cobalt chromium alloy and titanium alloy rods. The dynamic four-point bending test showed that force/displacement at a minimum force at 2,500,000 cycles was larger in the notch-free cobalt chromium alloy rod than in the notched cobalt chromium alloy rod. CONCLUSION: The notch-free curved cobalt chromium alloy rod is likely to maintain its curvature after spinal deformity surgery with a decreased risk of breakage and could overcome the problems of the conventional notched rod such as breakage and spring-back. LEVEL OF EVIDENCE: N/A.

来源:Spine (Phila Pa 1976),查看原文。版权归原作者所有。

引用

  1. Yamada K, Sudo H, Iwasaki N, et al. Mechanical Analysis of Notch-Free Pre-Bent Rods for Spinal Deformity Surgery. Spine (Phila Pa 1976). 2020. PMID: 31574057. doi:10.1097/BRS.0000000000003269

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。