特定骨水泥柄在移位性髋部骨折中术中骨折高发
回顾性队列研究比较Corail骨水泥柄(348例)与既往OmniFit EON骨水泥柄(77例)治疗移位性股骨颈骨折,Corail组术中股骨距或大转子骨折发生率为15.5%,对照组为2.7%(P < .05),无患者或术者因素解释该差异,结论认为应质疑假体设计。
要点
- Corail骨水泥柄组术中股骨距或大转子骨折发生率为15.5%,对照组为2.7%(P < .05)。
- 两组基线特征相似(P > .05),仅痴呆存在差异。
- 无患者相关或术者相关因素与Corail组骨折数增高相关(P > .05)。
核心结论
- Corail骨水泥柄在老年移位性股骨颈骨折中呈现异常数量的医源性术中骨折。
- 似乎无外部因素能解释该高骨折数,应质疑假体设计。
研究设计与样本
- 回顾性队列研究,比较历史队列(OmniFit EON骨水泥柄,77例)与新队列(Corail骨水泥柄,348例)的移位性股骨颈骨折患者。
- 由四名骨科医生审阅手术记录与术后X线片。
- 证据等级III。
主要发现
- Corail组术中股骨距或大转子骨折发生率为15.5%,对照组为2.7%(P < .05)。
- 两组基线特征相似(P > .05),仅痴呆存在差异。
- 无患者相关因素或术者相关因素与Corail组更高骨折数相关(P > .05)。
关键数值与推荐
- Corail组术中股骨距或大转子骨折发生率15.5%(348例中),对照组2.7%(77例中),P < .05。
局限性
- 原文未述局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: To reduce costs of orthopedic implants, the government decided to standardize implants used across different specialties in a group of hospitals located in the same geographic area. The usual cemented stem used in the context of intracapsular displaced geriatric hip fractures was replaced by another stem. Abnormal intraoperative calcar and trochanteric fractures were noted. The purpose of this study is to determine the incidence of intraoperative periprosthetic fractures following an intracapsular displaced hip fracture treated with this specific cemented stem compared to the previous implant. METHODS: This is a retrospective cohort study comparing an historic cohort of hip fractures treated with the OmniFit EON (Stryker, Kalamazoo, MI) cemented stem with a new cohort of patients who received the Corail (DePuy Synthes, Warsaw, IN) cemented stem. Four orthopedic surgeons reviewed operative reports and postoperative radiographs. RESULTS: The treatment group included 348 patients who received the Corail stem. The control group included 77 patients. The 2 groups had similar baseline characteristics (P > .05) except for the presence of dementia. Incidence of intraoperative calcar or greater trochanteric fracture was 15.5% for the Corail group and 2.7% for the control group (P < .05). No patient-related factors or surgeon-related factors were related to a higher number of fractures in the treatment group (P > .05). CONCLUSION: The Corail cemented stem presents an abnormal number of iatrogenic intraoperative fractures following displaced femoral neck fracture in our geriatric population. No external factor seems to explain this high number of fractures. Implant design should be questioned. LEVEL OF EVIDENCE: III.
引用
- Laflamme M, Angers M, Vachon J, et al. High Incidence of Intraoperative Fractures With a Specific Cemented Stem Following Intracapsular Displaced Hip Fracture. J Arthroplasty. 2019. PMID: 31594703. doi:10.1016/j.arth.2019.09.017
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。