髁限制型与旋转铰链型膝关节假体的生物力学差异数值研究
本研究用经验证的数值模型比较髁限制型(CCK)与旋转铰链型(RTH)假体在蹲起与弓步任务中的生物力学表现。RTH假体在胫骨柄尖端附近应力较高,CCK假体因立柱-凸轮机制使近端胫骨应力及假体微动增高。作者认为术者需考虑不同限制程度对骨应力与假体固定的影响。
要点
- RTH假体在胫骨柄尖端附近应力高于CCK假体。
- CCK假体因立柱-凸轮机制使近端胫骨应力高于RTH假体。
- CCK假体因扭转限制更大而假体微动高于RTH假体。
核心结论
- 翻修全膝关节置换中不同限制程度始终对应不同的生物力学输出。
- RTH假体以胫骨应力较高为特征,尤其在柄尖端附近区域。
- CCK假体因立柱-凸轮机制使近端胫骨应力及假体微动增高。
- 作者认为术者需考虑这些发现以保证患者最佳结局及不同限制程度引起的骨应力与假体固定变化。
研究设计与样本
- 采用经验证的数值模型分析CCK与RTH假体模型。
- 每种系统均在蹲起与弓步运动任务中进行分析。
- 分析并比较不同设计间的关节力、骨与假体应力及假体微动。
主要发现
- 不同活动在两种假体中诱导相似的骨应力分布。
- RTH假体相比CCK假体主要诱导较高应力,尤其在柄尖端附近区域。
- 在近端胫骨,CCK假体所得应力高于RTH设计计算值,归因于立柱-凸轮系统的存在。
- 相应地,髁限制型设计因扭转限制更大而显示更高的假体微动。
临床怎么用
- 作者认为术者需考虑这些发现以保证患者最佳结局及不同限制程度引起的骨应力与假体固定变化。
局限性
- 原文未报告局限性。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Different levels of constraint for total knee arthroplasty can be considered for revision surgeries. While prior studies have assessed the clinical impact and patient outcomes of condylar constrained knee (CCK) and rotating hinged (RTH) implants, nowadays little is known about the biomechanical effects induced by different levels of constraint on bone stress and implant micromotions. METHODS: CCK and RTH implant models were analyzed using a previously validated numerical model. Each system was investigated during a squat and a lunge motor task. The force in the joint, the bone and implant stresses, and micromotions in this latter were analyzed and compared among designs. RESULTS: Different activities induced similar bone stress distributions in both implants. The RTH implant induces mostly high stress compared to the CCK implant, especially in the region close to tip of the stem. However, in the proximal tibia, the stresses achieved with the CCK implant is higher than the one calculated for the RTH design, due to the presence of the post-cam system. Accordingly, the condylar constrained design shows higher implant micromotions due to the greater torsional constraint. CONCLUSION: Different levels of constraint in revision arthroplasty were always associated with different biomechanical outputs. RTH implants are characterized by higher tibial stress especially in the region close to the stem tip; condylar implants, instead, increase the proximal tibial stress and therefore implant micromotions, as a result of the presence of the post-cam mechanism. Surgeons will have to consider these findings to guarantee the best outcome for the patient and the related change in the bone stress and implant fixation induced by different levels of constrain in a total knee arthroplasty.
引用
- Andreani L, Pianigiani S, Bori E, et al. Analysis of Biomechanical Differences Between Condylar Constrained Knee and Rotating Hinged Implants: A Numerical Study. J Arthroplasty. 2019. PMID: 31473061. doi:10.1016/j.arth.2019.08.005
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。