文献速递关节外科

便携式加速度计导航系统在全髋关节置换术髋臼杯放置中的前瞻性随机对照研究

Portable Accelerometer-Based Navigation System for Cup Placement of Total Hip Arthroplasty: A Prospective, Randomized, Controlled Study.
PubMed 文献 · J Arthroplasty · 2019-08-27

前瞻性随机对照研究比较便携式加速度计导航系统与传统技术在全髋关节置换术髋臼杯定位中的效果,导航组55例、传统组55例。导航组髋臼外展角变异更小(P=0.001),偏离目标位置更少(P=0.001、0.016),位于Lewinnek安全区内的髋臼杯更多(51/55 vs 37/55,P=0.006),但手术时间平均延长10分钟。结论认为该导航系统可改善全髋关节置换术中的髋臼杯定位。

要点

  • 导航组髋臼外展角变异较传统组更小(P=0.001)。
  • 导航组偏离目标髋臼杯位置显著低于传统组(P=0.001、0.016)。
  • 导航组51/55髋臼杯位于Lewinnek安全区内,传统组37/55(P=0.006)。

核心结论

  • 使用便携式加速度计髋关节导航系统可改善全髋关节置换术中的髋臼杯定位。

研究设计与样本

  • 前瞻性随机临床研究,导航组55例、传统组55例,侧卧位后路入路行全髋关节置换术,术后通过骨盆X线片和CT扫描确定髋臼杯位置。

主要发现

  • 导航组平均髋臼外展角39.2°±4.6°(范围27°至50°),平均髋臼前倾角14.6°±6.1°(范围1°至27.5°)。
  • 传统组平均髋臼外展角42.9°±8.0°(范围23°至73°),平均髋臼前倾角11.6°±7.7°(范围-12.1°至25°)。
  • 导航组髋臼外展角变异更小(P=0.001),偏离目标髋臼杯位置显著低于传统组(P=0.001、0.016)。
  • 导航组51/55髋臼杯位于Lewinnek安全区内,传统组37/55(P=0.006)。
  • 导航手术时间较传统技术平均延长10分钟。

关键数值与推荐

  • 导航组51/55髋臼杯位于Lewinnek安全区内,传统组37/55(P=0.006)。
  • 导航组髋臼外展角变异更小(P=0.001),偏离目标髋臼杯位置显著低于传统组(P=0.001、0.016)。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Malposition of the acetabular component during total hip arthroplasty (THA) is associated with increased risk of dislocation, reduced range of motion, and accelerated wear. The purpose of this study is to compare cup positioning with a portable, accelerometer-based hip navigation system and conventional surgical technique. METHODS: In a prospective, randomized, clinical study, cups were implanted with a portable, accelerometer-based hip navigation system (navigation group; n = 55) or conventional technique (conventional group; n = 55). THA was conducted in the lateral position and through posterior approach. The cup position was determined postoperatively on pelvic radiograph and computed tomography scans. RESULTS: An average cup abduction of 39.2° ± 4.6° (range, 27° to 50°) and an average cup anteversion of 14.6° ± 6.1° (range, 1° to 27.5°) were found in the navigation group, and an average cup abduction of 42.9° ± 8.0° (range, 23° to 73°) and an average cup anteversion of 11.6° ± 7.7° (range, -12.1° to 25°) in the conventional group. A smaller variation in the navigation group was indicated for cup abduction (P = .001). The deviations from the target cup position were significantly lower in the navigation group (P = .001, .016). While only 37 of 55 cups in the conventional group were inside the Lewinnek safe zone, 51 of 55 cups in the navigation group were placed inside this safe zone (P = .006). The navigation procedure took a mean of 10 minutes longer than the conventional technique. CONCLUSION: Use of the portable, accelerometer-based hip navigation system can improve cup positioning in THA.