现代非骨水泥双动髋臼假体全髋置换10至16年结果
本研究回顾性评估310例初次全髋置换使用现代非骨水泥双动髋臼杯(Saturne)的影像学、临床和生存率结果,最终138髋完成术前术后分析,平均随访152.4个月。术后Harris与Postel-Merle d'Aubigne评分较术前显著提高(P<.001),无髋臼杯松动、骨溶解、脱位或感染,全队列10年生存率约98%。
要点
- 术后Harris与Postel-Merle d'Aubigne评分较术前显著提高(P<.001)。
- 末次随访无髋臼杯松动、无髋臼骨溶解、无假体或假体内脱位、无感染。
- 全队列(N=310)10年生存率约98%,3例因股骨骨折行股骨柄翻修。
核心结论
- 该非骨水泥双动髋臼杯在10年时无髋臼组件失败、无髋臼骨溶解,无假体或假体内脱位报告。
研究设计与样本
- 回顾性评估2001年4月至2005年12月单中心310例初次全髋置换,使用现代非骨水泥双动髋臼杯(Saturne)。
- 患者术后接受前瞻性临床与影像学随访,随访不足120个月的髋排除(5例失访、167例死亡)。
- 最终138髋纳入术前术后分析,平均随访152.4个月,行Kaplan-Meier生存分析。
主要发现
- 术后Harris与Postel-Merle d'Aubigne评分较术前显著提高(P<.001)。
- 末次随访无髋臼杯松动、无髋臼骨溶解。
- 无假体脱位、无假体内脱位、无感染报告。
- 全队列(N=310)全髋置换生存曲线10年约98%,3例因股骨骨折行股骨柄翻修。
- 报告1例腰大肌撞击。
局限性
- 本研究为回顾性评估。
- 单中心研究。
- 310例中167例死亡、5例失访,最终仅138髋纳入术前术后分析。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: The purpose of this study is to assess the radiographic results, clinical outcomes, and survivorship of a modern cementless dual-mobility cup (DMC) implant as a primary THA with a minimum of 10 years follow-up. METHODS: This study retrospectively assessed a series of 310 primary THAs using a modern-generation cementless DMC (Saturne acetabular cup) between April 2001 and December 2005 at a single center. Patients were followed prospectively clinically and radiographically after surgery. Hips with follow-up less than 120 months were excluded from the study (5 lost to follow-up and 167 deceased). In total, a cohort of 138 hips were included for preoperative and postoperative analysis with an average follow-up of 152.4 months. All complications were collected, and a Kaplan-Meier survival analysis was performed. RESULTS: There was a significant increase in the mean Harris and Postel-Merle d'Aubigne scores between preoperative and postoperative cohorts (P < .001). No loosening of the cup and no acetabular osteolysis were found at final follow-up. No prosthetic dislocation, no intraprosthetic dislocation, and no infections were reported. The survival curve of THA in the total cohort (N = 310) was about 98% at 10 years with 3 stem revisions for femoral fracture. One psoas impingement was also described. CONCLUSION: This study showed no acetabular component failure and no reported cases of acetabular osteolysis with this DMC acetabular component retention at 10 years. No prosthetic or intraprosthetic dislocation was reported.
引用
- Gaillard R, Kenney R, Delalande JL, et al. Ten- to 16-Year Results of a Modern Cementless Dual-Mobility Acetabular Implant in Primary Total Hip Arthroplasty. J Arthroplasty. 2019. PMID: 31353249. doi:10.1016/j.arth.2019.06.051
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。