新型机器人辅助全膝关节置换术的精确性:一项尸体研究

Accuracy of a New Robotically Assisted Technique for Total Knee Arthroplasty: A Cadaveric Study.
PubMed 文献 · J Arthroplasty · 2019-06-26

该尸体研究使用15具冰冻尸体标本(30膝)评估新型机器人辅助全膝关节置换系统(ROSA Knee System,Zimmer Biomet)的截骨精确性。8名经过培训的骨科医生采用同一机器人方案和3种不同假体设计完成手术,结果显示所有角度平均差异均小于1°,所有截骨厚度平均差异均小于0.7 mm,表明该系统能实现高精度的计划角度和截骨厚度。

要点

  • 所有角度平均差异均小于1°,标准差小于1°。
  • 所有截骨厚度平均差异小于0.7 mm,标准差小于1.1 mm。
  • 股骨屈曲角平均差异为0.95°,是唯一与0有显著差异的角度。

核心结论

  • 尽管本研究由Zimmer Biomet资助且仅使用Zimmer Biomet的假体、机器人和导航工具,但这项体外研究结果表明,医生使用该新型手术机器人进行全膝关节置换术能够实现高度精确的截骨,达到计划的角度和截骨厚度(以传统导航测量为准)。

研究设计与样本

  • 尸体研究,使用15具冰冻尸体标本(30膝),由8名经过培训的委员会认证骨科医生使用Zimmer Biomet膝关节、导航系统和机器人(ROSA Knee System;Zimmer Biomet)完成机器人辅助全膝关节置换术,采用同一机器人方案和3种不同假体设计。

主要发现

  • 所有角度平均差异均小于1°,标准差小于1°。
  • 6个角度中,除股骨屈曲角外,目标角度与测量值之间的平均差异与0无显著差异;股骨屈曲角平均差异为0.95°。
  • 髋-膝-踝轴平均差异为-0.03° ± 0.87°。
  • 所有截骨厚度平均差异均小于0.7 mm,标准差小于1.1 mm。

局限性

  • 本研究由Zimmer Biomet资助。
  • 本研究仅使用了Zimmer Biomet的假体、机器人和导航工具。
  • 本研究为体外研究。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Although the utility of robotic surgery has already been proven in cadaveric studies, it is our hypothesis that this newly designed robotically assisted system will achieve a high level of accuracy for bone resection. Therefore, we aimed to analyze in a cadaveric study the accuracy to achieve targeted angles and resection thickness. METHODS: For this study, 15 frozen cadaveric specimens (30 knees) were used. In this study, Zimmer Biomet (Warsaw, IN) knees, navigation system, and robot (ROSA Knee System; Zimmer Biomet) were used. Eight trained, board-certified orthopedic surgeons performed robotically assisted total knee arthroplasty implantation using the same robotic protocol with 3 different implant designs. The target angles obtained from the intraoperative planning were then compared to the angles of the bone cuts performed using the robotic system and measured with the computer-assisted system considered to be the gold standard. For each bone cut the resection thickness was measured 3 times by 2 different observers and compared to the values for the planned resections. RESULTS: All angle mean differences were below 1° and standard deviations below 1°. For all 6 angles, the mean differences between the target angle and the measured values were not significantly different from 0 except for the femoral flexion angle which had a mean difference of 0.95°. The mean hip-knee-ankle axis difference was -0.03° ± 0.87°. All resection mean differences were below 0.7 mm and standard deviations below 1.1mm. CONCLUSION: Despite the fact that this study was funded by Zimmer Biomet and only used Zimmer Biomet implants, robot, and navigation tools, the results of our in vitro study demonstrated that surgeons using this new surgical robot in total knee arthroplasty can perform highly accurate bone cuts to achieve the planned angles and resection thickness as measured using conventional navigation.