电灼损伤可降低植入物疲劳强度:病例报告与体外研究

Electrocautery Damage Can Reduce Implant Fatigue Strength: Cases and in Vitro Investigation.
PubMed 文献 · J Bone Joint Surg Am · 2019-01-01

本文报告2例全髋关节置换翻修术中陶瓷部件碎裂后股骨柄失效的临床病例,并对电灼设备接触后的钛合金股骨柄进行体外循环加载至失效测试及材料微观结构分析。结果显示电灼损伤使体外失效载荷较未损伤标本降低达47%,材料微观结构改变延伸约2.7 mm、深度550 µm,分3个结构区。作者认为翻修术中应避免电灼设备接触股骨部件。

要点

  • 翻修术中电灼设备接触植入物可改变材料微观结构并可能降低疲劳强度。
  • 体外测试中电灼损伤使股骨颈标本失效载荷较未损伤标本降低达47%。
  • 取回部件表面发现变色和熔化痕迹,包括裂纹起始处,并检出电灼尖端材料转移。

核心结论

  • 翻修术中电灼设备接触植入物不能总是避免,但基于本研究发现,微观结构改变和植入物疲劳强度潜在降低使植入物失效风险增加。
  • 作者认为外科医生和电灼设备制造商应关注这一问题。

研究设计与样本

  • 报告2例全髋关节置换翻修术中陶瓷部件碎裂后股骨柄失效的临床病例,并对取回部件进行失效原因分析。
  • 对钛合金股骨柄在2个不同位置(基底部和股骨颈中部附近)施加电灼后进行体外阶梯增加载荷至植入物断裂的循环加载至失效测试。
  • 对电灼标记周围局部材料结构进行详细表征。

主要发现

  • 取回部件上发现表面变色和熔化痕迹,包括前外侧区域裂纹起始处,可能降低了材料疲劳强度。
  • 元素分析提示电灼尖端存在材料转移。
  • 电灼设备造成的表面损伤使体外失效载荷较未损伤股骨颈标本降低达47%。
  • 材料分析显示微观结构发生相关改变,延伸约2.7 mm、深度550 µm,可分为3个结构区。

关键数值与推荐

  • 电灼损伤使体外失效载荷较未损伤股骨颈标本降低达47%。
  • 材料微观结构改变延伸约2.7 mm、深度550 µm,可分为3个结构区。

临床怎么用

  • 作者认为,翻修术中应避免电灼设备与股骨部件接触,以减少后续股骨颈骨折的机会。
  • 作者认为外科医生和电灼设备制造商应关注电灼设备接触植入物的问题。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The risk of femoral stem fracture after total hip replacement is low and can often be associated with a specific implant system or other factors that may reduce the fatigue strength. Additionally, damage to a metal component during revision surgery by an electrocautery device may further affect the fatigue behavior. METHODS: Two clinical cases of stem failure after revision of fractured ceramic components are presented; the retrieved components were analyzed for the cause of failure. In vitro cyclic load-to-failure testing of titanium alloy femoral stems after electrocautery application at 2 different locations (at the base and about midway on the femoral neck) was performed using a stepwise increase in load until implant fracture occurred. In addition, a detailed characterization of the local material structure around the electrocautery marks was performed. RESULTS: Superficial discoloration and melting marks were found on the retrieved components, including at the location of crack initiation in the anterolateral region, which may have reduced the fatigue strength of the material. In addition, elemental analysis indicated material transfer from the electrocautery tip. Damage to the surface by the electrocautery device significantly reduced the in vitro load to failure by up to 47% compared with that of undamaged femoral neck specimens. Material analysis revealed a relevant modification in microstructure, with an extension of approximately 2.7 mm and a depth of 550 µm, which could be divided in 3 structural zones. CONCLUSIONS: Intraoperative electrocautery device contact with the implant during surgical revision of a total hip replacement cannot always be avoided. However, on the basis of our findings, the risk of implant failure is increased due to a change in microstructure and a potential reduction of the implant's fatigue strength. Surgeons and manufacturers of electrocautery devices should be aware of this concern. CLINICAL RELEVANCE: During revision surgery, contact between an electrocautery device and the femoral component should be avoided to reduce the chance of subsequent femoral neck fracture.