一种常用双模块全髋关节置换设计中股骨颈骨折的相关因素:斯洛文尼亚全国多中心研究

What Factors Are Associated With Neck Fracture in One Commonly Used Bimodular THA Design? A Multicenter, Nationwide Study in Slovenia.
PubMed 文献 · Clin Orthop Relat Res · 2019-01-01

该回顾性全国多中心研究分析了2767例Profemur® Z双模块股骨柄,平均随访8年,无菌性松动12年生存率为97%(95%CI ±1%),模块股骨颈骨折发生率为0.83%(23例),植入时间、长颈、钴铬合金颈、年轻年龄和男性与颈骨折相关。作者认为该双模块柄骨折率不可接受,尤其年轻男性钴铬长颈患者。

要点

  • Profemur® Z双模块股骨柄12年无菌性松动生存率为97%(95%CI ±1%)。
  • 模块股骨颈骨折发生率为0.83%(23/2767),其中20例为男性,19例为长颈。
  • 长颈(OR 6.77)、钴铬合金颈(OR 5.7)、男性(OR 3.98)和年轻年龄(OR 0.91)与颈骨折相关。

核心结论

  • Profemur® Z柄的松动率和颈骨折率低于一些既往发表的系列。
  • 在初次全髋关节置换中使用模块股骨颈增加颈骨折风险,尤其是年轻男性钴铬长颈患者。
  • 作者认为所分析的双模块柄骨折率不可接受,尤其在年轻男性患者中。
  • 作者认为对大多数患者,使用该器械的风险大于获益,数十例患者经历了若使用不同股骨柄则不会发生的翻修和并发症。

研究设计与样本

  • 回顾性、全国性、多中心研究,回顾斯洛文尼亚四家医院2002年至2015年间植入的2767例Profemur® Z双模块股骨柄,平均随访8年(范围3天至15年)。
  • 四家医院期间共行26132例初次全髋关节置换,该植入物用于其中2767例(11%)。
  • 随访至死亡、翻修或2018年3月1日随访结束。
  • 女性植入1438例(52%),男性1329例(48%);钛合金颈2489例(90%),钴铬合金颈278例(10%);手术时平均BMI为29 kg/m(SD ±5 kg/m)。

主要发现

  • 无菌性股骨柄翻修55例(2%);12年无菌性松动生存率为97%(95%CI ±1%)。
  • 模块股骨颈骨折发生于23例患者(0.83%),平均BMI为29 kg/m(SD ±4 kg/m);20例为男性,23个骨折颈中19个为长颈。
  • 植入时间(OR 0.55;95%CI 0.46-0.66;p < 0.001)、长颈(OR 6.77;95%CI 2.1-22.2;p = 0.002)、钴铬合金颈(OR 5.7;95%CI 1.6-21.1;p = 0.008)、年轻年龄(OR 0.91;95%CI 0.86-0.96;p < 0.001)和男性(OR 3.98;95%CI 1.04-14.55;p = 0.043)与颈骨折相关。

关键数值与推荐

  • 颈骨折相关因素:植入时间OR 0.55(95%CI 0.46-0.66;p < 0.001);长颈OR 6.77(95%CI 2.1-22.2;p = 0.002);钴铬合金颈OR 5.7(95%CI 1.6-21.1;p = 0.008);年龄OR 0.91(95%CI 0.86-0.96;p < 0.001);男性OR 3.98(95%CI 1.04-14.55;p = 0.043)。
  • 12年无菌性松动生存率97%(95%CI ±1%)。
  • 模块股骨颈骨折发生率0.83%(23例)。

临床怎么用

  • 作者认为在初次全髋关节置换中使用模块股骨颈增加颈骨折风险,尤其是年轻男性钴铬长颈患者。
  • 作者认为对大多数患者,使用该双模块柄的风险大于获益。

局限性

  • 原文未报告局限性

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Femoral stems with bimodular (head-neck as well as neck-body) junctions were designed to help surgeons address patients' hip anatomy individually. However, arthroplasty registers have reported higher revision rates in stems with bimodular junctions than in stems with modularity limited to the head-neck trunnion. However, to our knowledge, no epidemiologic study has identified patient-specific risk factors for modular femoral neck fractures, and some stems using these designs still are produced and marketed. QUESTIONS/PURPOSES: The purposes of this study were (1) to establish the survival rate free from aseptic loosening of one widely used bimodular THA design; (2) to define the proportion of patients who experienced a fracture of the stem's modular femoral neck; and (3) to determine factors associated with neck fracture. METHODS: In this retrospective, nationwide, multicenter study, we reviewed 2767 bimodular Profemur® Z stems from four hospitals in Slovenia with a mean followup of 8 years (range, 3 days to 15 years). Between 2002 and 2015, the four participating hospitals performed 26,132 primary THAs; this implant was used in 2767 of them (11%). The general indications for using this implant were primary osteoarthritis (OA) in 2198 (79%) hips and other indications in 569 (21%) hips. We followed patients from the date of the index operation to the date of death, date of revision, or the end of followup on March 1, 2018. We believe that all revisions would be captured in our sample, except for patients who may have emigrated outside the country, but the proportion of people immigrating to Slovenia is higher than the proportion of those emigrating from it; however, no formal accounting for loss to followup is possible in a study of this design. There were 1438 (52%) stems implanted in female and 1329 (48%) in male patients, respectively. A titanium alloy neck was used in 2489 hips (90%) and a cobalt-chromium neck in 278 (10%) hips. The mean body mass index (BMI) at the time of operation was 29 kg/m (SD ± 5 kg/m). We used Kaplan-Meier analysis to establish survival rates, and we performed a chart review to determine the proportion of patients who experienced femoral neck fractures. A binary logistic regression model that controlled for the potential confounding variables of age, sex, BMI, time since implantation, type of bearing, diagnosis, hospital, neck length, and neck material was used to analyze neck fractures. RESULTS: There were 55 (2%) aseptic stem revisions. Survival rate free from aseptic loosening at 12 years was 97% (95% confidence interval [CI] ± 1%). Fracture of the modular neck occurred in 23 patients (0.83%) with a mean BMI of 29 kg/m (SD ± 4 kg/m.) Twenty patients with neck fractures were males and 19 of 23 fractured necks were long. Time since implantation (odds ratio [OR], 0.55; 95% CI 0.46-0.66; p < 0.001), a long neck (OR, 6.77; 95% CI, 2.1-22.2; p = 0.002), a cobalt-chromium alloy neck (OR, 5.7; 95% CI, 1.6-21.1; p = 0.008), younger age (OR, 0.91; 95% CI, 0.86-0.96; p < 0.001), and male sex (OR, 3.98; 95% CI, 1.04-14.55; p = 0.043) were factors associated with neck fracture. CONCLUSIONS: The loosening and neck fracture rates of the Profemur® Z stem were lower than in some of previously published series. However, the use of modular femoral necks in primary THA increases the risk for neck fracture, particularly in young male patients with cobalt-chromium long femoral necks. The bimodular stem we analyzed fractured unacceptably often, especially in younger male patients. For most patients, the risks of using this device outweigh the benefits, and several dozen patients had revisions and complications they would not have had if a different stem had been used. LEVEL OF EVIDENCE: Level III, therapeutic study.