两种高交联聚乙烯热处理方式体内性能比较:髋臼内衬重熔与退火对比

Comparing In Vivo Performance of Two Highly Cross-Linked Polyethylene Thermal Treatments: Remelting vs Annealing in Acetabular Liners.
PubMed 文献 · J Arthroplasty · 2019-03-14

回顾性研究比较重熔A-CLASS(MicroPort)高交联聚乙烯内衬与顺序退火高交联聚乙烯内衬的磨损率及翻修情况,共纳入重熔组80例、退火组53例进行磨损分析。两组稳态线性及体积磨损率无显著差异,均低于0.1 mm/y线性磨损阈值,无因内衬相关并发症翻修。作者认为重熔内衬磨损率与顺序退火内衬相当,需进一步长期研究确保抗疲劳和体内抗氧化性能。

要点

  • 重熔与退火高交联聚乙烯内衬的稳态线性磨损率无显著差异,均低于0.1 mm/y阈值。
  • 两组稳态体积磨损率无显著差异,无因内衬相关并发症翻修。
  • 股骨头尺寸≥36 mm与32、28 mm组磨损率无显著差异,髋臼杯外倾>50°与标准位置组磨损率也无显著差异。

核心结论

  • 重熔A-CLASS高交联聚乙烯髋臼内衬的磨损率与顺序退火高交联聚乙烯内衬相当,需进一步长期研究以确保可接受的抗疲劳性和体内抗氧化性。

研究设计与样本

  • 回顾性研究,使用Martell Hip Analysis Suite评估线性及体积磨损率,并通过翻修手术发生率评估临床性能。
  • 磨损分析共纳入80个重熔内衬和53个退火内衬,所有髋关节均进行翻修审查。

主要发现

  • 重熔组与退火组稳态线性磨损率分别为0.01(-0.07至0.14)mm/y与-0.01(-0.11至0.1)mm/y,差异无统计学意义(P=0.28)。
  • 重熔组与退火组稳态体积磨损率分别为-1.03(-30.99至45.43)mm³/y与-1.31(-32.23至23.70)mm³/y,差异无统计学意义(P=0.30)。
  • 两组队列线性磨损率均低于0.1 mm/y阈值。
  • 股骨头尺寸≥36 mm患者与32及28 mm股骨头尺寸患者的磨损率无显著差异(P=0.60)。
  • 髋臼假体过度垂直(>50°)患者与标准髋臼假体方位患者的磨损率无显著差异(P=0.97)。

关键数值与推荐

  • 稳态线性磨损率:重熔组0.01(-0.07至0.14)mm/y vs 退火组-0.01(-0.11至0.1)mm/y(P=0.28)。
  • 稳态体积磨损率:重熔组-1.03(-30.99至45.43)mm³/y vs 退火组-1.31(-32.23至23.70)mm³/y(P=0.30)。
  • 线性磨损阈值:两组均低于0.1 mm/y。
  • 股骨头尺寸≥36 mm vs 32及28 mm的磨损率比较P=0.60。
  • 髋臼假体过度垂直(>50°)vs 标准方位的磨损率比较P=0.97。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文结论指出需进一步长期研究以确保可接受的抗疲劳性和体内抗氧化性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: The introduction of highly cross-linked polyethylene (HXLPE) acetabular liners has greatly improved the wear performance of metal-on-PE bearing surfaces used in total hip arthroplasty. Changing the sterilization environment and adding thermal treatments, such as remelting or annealing, were introduced to limit on-shelf and in vivo oxidation of cross-linked liners. This study compares the wear properties of the remelted A-CLASS (MicroPort) HXLPE liner to a sequentially annealed HXLPE. METHODS: This retrospective study assessed linear and volumetric wear rates using Martell Hip Analysis Suite, and clinical performance through incidences of revision surgeries. A total of 80 remelted and 53 annealed liners were included in the wear analysis. All hips were reviewed for revisions. RESULTS: There were no significant differences in steady-state linear or volumetric wear rates for remelted and annealed liners, 0.01 (-0.07 to 0.14) vs -0.01 (-0.11 to 0.1) mm/y (P = .28) and -1.03 (-30.99 to 45.43) vs -1.31 (-32.23 to 23.70) mm3/y (P = .30), respectively. Both cohorts were below the 0.1 mm/y linear wear threshold. The wear rates for patients with femoral head sizes ≥36 mm were not significantly different than those with 32 and 28 mm femoral head sizes (P = .60). Similarly, wear rates for patients with an excessively vertical acetabular component (>50°) were not significantly different than those with standard acetabular component orientations (P = .97). No hips were revised due to liner-related complications. CONCLUSION: The wear rates of the A-CLASS remelted HXLPE acetabular liner wear rates were comparable to those of a sequentially annealed HXLPE. Further long-term studies are required to ensure acceptable resistance to fatigue and in vivo oxidation.