使用模块化股骨锥形连接件的初次全髋关节置换术在至少5年随访时失败率高
研究回顾了2002至2004年间一位术者使用ALFA II模块化股骨柄进行的221例初次全髋关节置换术,在平均6.5年随访时全因生存率为81%(95%CI=69-90),25例(19.4%)需翻修,其中52.0%为模块组件分离,32.0%为假体骨折;体重指数(OR=1.080)和偏心距(OR=1.254)是机械性失败的独立危险因素。作者警告不要使用类似设计的模块化柄进行初次全髋关节置换。
要点
- 平均6.5年随访时模块化柄全因生存率为81%(95%CI=69-90)。
- 25例(19.4%)需翻修,52.0%因模块组件分离,32.0%因假体骨折。
- 体重指数(OR=1.080)和偏心距(OR=1.254)是机械性失败的独立危险因素。
核心结论
- 研究的模块化股骨柄髋系统在中远期随访中显示高失败率,作者警告不要使用类似设计进行初次全髋关节置换。
研究设计与样本
- 连续系列回顾性研究,纳入2002年至2004年间一位术者使用ALFA II模块化股骨柄进行的221例初次全髋关节置换术患者,术后至少随访5年评估生存率。
主要发现
- 221例患者中,28例(12.7%)在充分随访前因与手术无关的原因死亡,64例(29.0%)失访。
- 剩余129例患者平均随访6.5年(范围5-8年),模块化柄系统的全因生存率为81%(95%CI=69-90)。
- 25例(19.4%)需翻修手术,其中52.0%因模块组件分离,32.0%因假体骨折,12.0%因不稳定/多次脱位,4.0%因慢性感染性全髋关节置换。
- 体重指数(OR=1.080)和偏心距(OR=1.254)是模块化柄系统机械性失败的独立危险因素。
关键数值与推荐
- 全因生存率81%(95%CI=69-90)对应平均6.5年随访的129例患者。
- 翻修率19.4%(25/129例),其中模块组件分离占52.0%,假体骨折占32.0%。
- 体重指数OR=1.080,偏心距OR=1.254,为机械性失败的独立危险因素。
临床怎么用
- 作者警告不要使用类似设计的模块化柄进行初次全髋关节置换。
局限性
- 28例(12.7%)在充分随访前因与手术无关的原因死亡,64例(29.0%)失访。
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
BACKGROUND: Modular femoral trunnions enable the surgeon to independently adjust offset, leg length, and anteversion in total hip arthroplasty (THA). However, modularity may result in an increased risk of fretting and corrosion along with a higher risk of implant dissociation or fracture. The purpose of this study is to evaluate mid-term survivorship of THAs using a cementless modular system. METHODS: A consecutive series of 221 patients who underwent a primary THA using the ALFA II modular stem by a single surgeon between 2002 and 2004 were reviewed. Survivorship of the ALFA II modular hip system was evaluated at a minimum of 5 years postoperatively. RESULTS: Of the 221 patients, 28 (12.7%) died from causes unrelated to the surgery before adequate follow-up, and 64 (29.0%) patients were lost to follow-up. The remaining 129 patients had a mean 6.5-year (range: 5-8 years) follow-up. All-cause survivorship of the modular stem system was 81% (95% confidence interval = 69-90) at a mean 6.5-year follow-up. Of the 25 (19.4%) cases requiring revision surgery, 52.0% was for dissociation of the modular components, 32.0% was for fracture of the prosthesis, 12.0% was for instability/multiple dislocations, and 4.0% was for chronic septic THA. Body mass index (odds ratio = 1.080) and offset (odds ratio = 1.254) were independent risk factors for mechanical failures of the modular stem system. CONCLUSION: The modular stem hip system of interest in this study demonstrates a high failure rate at mid-term follow-up, and we caution against the use of similar designs in primary THAs.
引用
- Nahhas CR, Yi PH, Moric M, et al. High Failure at a Minimum 5-Year Follow-Up in Primary Total Hip Arthroplasty Using a Modular Femoral Trunnion. J Arthroplasty. 2019. PMID: 30982763. doi:10.1016/j.arth.2019.03.033