半球形多孔涂层髋臼组件联合近端羟基磷灰石涂层解剖型股骨组件的疗效:23至26年随访更新
本研究对既往报道的Duraloc-Profile非骨水泥全髋关节置换系统队列进行23至26年随访更新,73例患者(82髋)平均年龄56.8岁。所有髋臼和股骨组件均固定良好并显示骨长入,25年无组件翻修概率为83.2%(95%CI 74.5至91.8),结论认为该系统长期骨固定效果优良,但建议监测磨损和晚期脱位。
要点
- 23至26年随访时所有髋臼和股骨组件均固定良好并显示骨长入迹象。
- 25年无组件翻修概率为83.2%(95%CI 74.5至91.8;风险人数41)。
- 股骨平均穿透15年为1.52 mm(sd 0.8),25年为1.92 mm(sd 1.2)。
核心结论
- Duraloc-Profile全髋关节置换系统显示出优良的长期骨固定。
- 建议对该人群监测磨损和晚期脱位,该人群手术时相对年轻。
研究设计与样本
- 对既往报道的队列进行23至26年随访更新,纳入99例原始患者(111髋)中73例(82髋)进行临床和影像学评估。
- 73例患者平均年龄56.8岁(21至70岁),女性40例(44髋),男性33例(38髋),最短随访23年。
主要发现
- 所有髋臼和股骨组件均固定良好并显示骨长入迹象。
- 9个髋臼组件因磨损-骨溶解相关问题翻修,4个因晚期脱位翻修。
- 25年无组件翻修概率为83.2%(95%CI 74.5至91.8;风险人数41)。
- 10髋观察到髋臼骨溶解;股骨平均穿透15年为1.52 mm(sd 0.8),25年为1.92 mm(sd 1.2)。
- ROC分析显示股骨平均穿透≥0.11 mm/年与骨溶解出现相关;25年Kaplan-Meier生存率:髋臼骨溶解89.9%(95%CI 83.3至96.5),近端股骨骨溶解92.1%(95%CI 86.1至98.2),股骨骨量减少75.5%(95%CI 66.5至84.5)。
关键数值与推荐
- 25年无组件翻修概率83.2%(95%CI 74.5至91.8)。
- 25年Kaplan-Meier生存率:髋臼骨溶解89.9%(95%CI 83.3至96.5),近端股骨骨溶解92.1%(95%CI 86.1至98.2),股骨骨量减少75.5%(95%CI 66.5至84.5)。
- 股骨平均穿透≥0.11 mm/年与骨溶解出现相关。
临床怎么用
- 作者建议对该人群监测磨损和晚期脱位。
局限性
- 原文未报告局限性
以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。
原文
AIMS: We previously reported the long-term results of the cementless Duraloc-Profile total hip arthroplasty (THA) system in a 12- to 15-year follow-up study. In this paper, we provide an update on the clinical and radiological results of a previously reported cohort of patients at 23 to 26 years´ follow-up. PATIENTS AND METHODS: Of the 99 original patients (111 hips), 73 patients (82 hips) with a mean age of 56.8 years (21 to 70) were available for clinical and radiological study at a minimum follow-up of 23 years. There were 40 female patients (44 hips) and 33 male patients (38 hips). RESULTS: All acetabular and femoral components were well fixed and showed signs of bone ingrowth. Nine acetabular components were revised due to wear-osteolysis-related problems and four due to late dislocation. The probability of not having component revision at 25 years was 83.2% (95% confidence interval (CI) 74.5 to 91.8; number at risk 41). Acetabular osteolysis was observed in ten hips. The mean femoral head penetration was 1.52 mm (sd 0.8) at 15 years and 1.92 mm (sd 1.2) at 25 years. Receiver operating characteristic (ROC) analysis revealed that mean femoral penetration with a value of 0.11 mm/year or more was associated with the appearance of osteolysis. The 25-year Kaplan-Meier survival with different endpoints was 89.9% for acetabular osteolysis (95% CI 83.3 to 96.5), 92.1% for proximal femoral osteolysis (95% CI 86.1 to 98.2), and 75.5% for femoral osteopenia (95% CI 66.5 to 84.5). CONCLUSION: The Duraloc-Profile THA system showed excellent long-term bone fixation. Nevertheless, monitoring is recommended in order to detect wear and late dislocations in this population that was relatively young at the time of surgery. Cite this article: Bone Joint J 2019;101-B:378-385.
引用
- García-Rey E, Carbonell-Escobar R, Cordero-Ampuero J, et al. Outcome of a hemispherical porous-coated acetabular component with a proximally hydroxyapatite-coated anatomical femoral component: an update at 23 to 26 years’ follow-up. Bone Joint J. 2019. PMID: 30929484. doi:10.1302/0301-620X.101B4.BJJ-2018-1223.R1
供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。