患者特异性器械不影响全膝关节置换术中股骨假体旋转对线与围手术期失血:一项前瞻性随机对照试验

Patient-Specific Instrumentation Does Not Affect Rotational Alignment of the Femoral Component and Perioperative Blood Loss in Total Knee Arthroplasty: A Prospective, Randomized, Controlled Trial.
PubMed 文献 · J Arthroplasty · 2019-03-13

该前瞻性随机对照试验将69例连续入组的全膝关节置换术患者随机分为患者特异性器械(PSI)组与常规器械对照组,术后2个月行膝关节CT测量股骨假体相对经上髁轴的旋转与胫骨假体后倾。结果显示PSI未显著改善股骨假体旋转对线,也未减少离群值,胫骨后倾、失血、总手术时间与缺血时间均无显著改善,而PSI组胫骨再截骨次数显著高于对照组(P=0.0003)。

要点

  • PSI未显著改善全膝关节置换术中股骨假体旋转对线,也未减少离群值。
  • PSI组胫骨再截骨次数显著高于对照组(P=0.0003)。
  • 胫骨后倾、失血、总手术时间与缺血时间在两组间均无显著改善。

核心结论

  • PSI与常规器械相比未改善全膝关节置换术中股骨假体旋转对线的准确性,也未影响本研究的其他次要观察变量。
  • 作者认为本研究结果不支持在标准全膝关节置换术中常规使用PSI。

研究设计与样本

  • 前瞻性、随机、对照临床试验,证据等级为Level I。
  • 共评估133例连续患者是否符合条件,最终入组69例,采用区组随机化分为PSI组与对照组。

主要发现

  • PSI组与对照组相比,股骨假体旋转对线无显著改善,离群值数量也无减少。
  • 胫骨后倾、失血、总手术时间与缺血时间均未发现显著改善。
  • PSI组所需胫骨再截骨次数显著高于对照组(P=0.0003)。

关键数值与推荐

  • 股骨假体旋转以术后2个月膝关节CT测量,参照经上髁轴。
  • 胫骨假体后倾同样经术后2个月膝关节CT测量。
  • 住院期间记录手术时间,并计算总血容量丢失与估计红细胞丢失。

临床怎么用

  • 原文结论为研究结果不支持在标准全膝关节置换术中常规使用PSI。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Proposed aims of patient-specific instrumentation (PSI) for total knee arthroplasty (TKA) are to improve accuracy of component alignment, while reducing blood loss and surgical time. The primary goal of this prospective, randomized, controlled, clinical trial is to verify whether PSI improves the rotational alignment of the femoral component in comparison to conventionally implanted TKA. METHODS: One-hundred thirty-three consecutive patients were assessed for eligibility. Block randomization was performed to allocated patients in the treatment (PSI) or control group. During hospital stay, surgical times were recorded, and total blood volume loss and estimated red blood cell were calculated. Two months after surgery, a computed tomography of the knee was obtained to measure femoral component rotation to the transepicondylar axis and tibial component slope. RESULTS: Sixty-nine patients were enrolled. PSI did neither result in a significant improvement in femoral component rotation nor result in a reduction of outliers, as compared with conventional instrumentation. No significant improvement in terms of tibial slope, blood loss, total surgical time, and ischemia time could be identified. The number of tibial recuts required in the PSI group was significantly higher than in the control group (P = .0003). CONCLUSION: PSI does not improve the accuracy of femoral component rotation in TKA in comparison to conventional instrumentation. Moreover, PSI did not appear to influence any of the other variables investigated as secondary goals by this study. The results of this study do not support its routine use during standard TKA. LEVEL OF EVIDENCE: Level I, randomized, controlled trial.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Randelli PS, Menon A, Pasqualotto S, et al. Patient-Specific Instrumentation Does Not Affect Rotational Alignment of the Femoral Component and Perioperative Blood Loss in Total Knee Arthroplasty: A Prospective, Randomized, Controlled Trial. J Arthroplasty. 2019. PMID: 30979672. doi:10.1016/j.arth.2019.03.018