文献速递关节外科

锥度设计是否影响陶瓷对聚乙烯全髋关节置换术中的锥度微动腐蚀?一项取回物分析

Does Taper Design Affect Taper Fretting Corrosion in Ceramic-on-Polyethylene Total Hip Arthroplasty? A Retrieval Analysis.
PubMed 文献 · J Arthroplasty · 2019-03-06

该取回物分析纳入2002年至2017年120例氧化锆增韧氧化铝股骨头的陶瓷对聚乙烯全髋关节置换系统,评估11/13、12/14、16/18和V40四种锥度设计的微动、腐蚀与损伤。结果显示V40与16/18锥度的微动和腐蚀合计评分显著高于12/14锥度,植入时长与微动、腐蚀均呈弱正相关。

要点

  • V40与16/18锥度的微动和腐蚀合计评分显著高于12/14锥度(均P≤.001)。
  • 植入时长与锥度微动(ρ=0.224,P=.016)和腐蚀(ρ=0.253,P=.006)呈弱正相关。
  • 16/18锥度组患者体重指数高于V40锥度组(P=.012)。

核心结论

  • 陶瓷对聚乙烯全髋关节置换植入物中观察到锥度微动和腐蚀,V40与16/18锥度最重、12/14锥度最轻。
  • 锥度设计特征的差异可能导致锥度-股骨头界面微动增加,从而引起微动和腐蚀增加。

研究设计与样本

  • 经机构审查委员会批准,查询2002年至2017年植入物取回库,识别出120例带氧化锆增韧氧化铝股骨头的取回陶瓷对聚乙烯全髋关节置换系统。
  • 对股骨柄锥部进行目视评估并分级锥度界面的微动、腐蚀和损伤,查阅病历获取患者人口学与植入物特征。
  • 采用Spearman相关与秩和检验(Dunn事后检验)进行统计分析,显著性水平α=0.05。

主要发现

  • 共评估四种锥度设计:11/13(n=18)、12/14(n=53)、16/18(n=21)、V40(n=28)。
  • 各锥度组间在植入时长、侧别、患者年龄和性别上无统计学显著人口学差异,但16/18锥度组患者体重指数高于V40锥度组(P=.012)。
  • 植入时长与锥部微动(ρ=0.224,P=.016)和腐蚀(ρ=0.253,P=.006)均呈弱正相关。
  • V40与16/18锥度的微动和腐蚀合计评分显著高于12/14锥度(均P≤.001)。

关键数值与推荐

  • 四种锥度设计及例数:11/13(n=18)、12/14(n=53)、16/18(n=21)、V40(n=28)。
  • 植入时长与微动相关ρ=0.224(P=.016),与腐蚀相关ρ=0.253(P=.006)。
  • V40与16/18锥度的微动和腐蚀合计评分显著高于12/14锥度,均P≤.001。
  • 16/18锥度组体重指数高于V40锥度组,P=.012。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Ceramic-on-polyethylene (CoP) implants have exhibited lower fretting and corrosion scores than metal-on-polyethylene implants. This study aims at investigating the effect of taper design on taper corrosion and fretting in modular CoP total hip arthroplasty (THA) systems. METHODS: Under an institutional review board--approved protocol, a query of an implant retrieval library from 2002 to 2017 identified 120 retrieved CoP THA systems with zirconia toughened alumina femoral heads. Femoral stem trunnions were visually evaluated and graded for fretting, corrosion, and damage at the taper interface. Medical records were reviewed for patient demographics and implant characteristics. Data were statistically analyzed using Spearman correlation and rank-sum tests with a Dunn's post hoc test, with a significance level of α = 0.05. RESULTS: Four different taper designs were evaluated: 11/13 (n = 18), 12/14 (n = 53), 16/18 (n = 21), and V40 (n = 28). There were no statistically significant demographic differences between taper groups for duration of implantation, laterality, patient age, and patient sex, but patients with 16/18 tapers had a higher body mass index than V40 tapers (P = .012). Duration of implantation had a weak positive correlation with both trunnion fretting (ρ = 0.224, P = .016) and corrosion (ρ = 0.253, P = .006). Summed fretting and corrosion scores were significantly greater on the V40 and 16/18 tapers compared with the 12/14 tapers (all P ≤ .001). CONCLUSION: Taper fretting and corrosion were observed in CoP THA implants and were greatest with V40 and 16/18 tapers and lowest with 12/14 tapers. Differences in taper design characteristics may lead to greater micromotion at the taper-head interface, leading to increased fretting and corrosion.

来源:J Arthroplasty,查看原文。版权归原作者所有。

引用

  1. Siljander MP, Gehrke CK, Wheeler SD, et al. Does Taper Design Affect Taper Fretting Corrosion in Ceramic-on-Polyethylene Total Hip Arthroplasty? A Retrieval Analysis. J Arthroplasty. 2019. PMID: 31000401. doi:10.1016/j.arth.2019.02.058

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。