文献速递关节外科

氧化锆股骨头能否减少全髋关节置换术中的摩擦腐蚀?对取回部件的比较研究

Do oxidized zirconium heads decrease tribocorrosion in total hip arthroplasty? A study of retrieved components.
PubMed 文献 · Bone Joint J · 2019-01-01

该研究对翻修全髋关节置换术中取回的28枚氧化锆(OxZr)股骨头与28枚按人口学匹配的钴铬(CoCr)股骨头进行对比,采用改良Goldberg评分法评估female taper的微动与腐蚀。结果显示OxZr组的平均腐蚀评分和微动评分均显著低于CoCr组,OxZr股骨头近端区域损伤较远端更重,CoCr组中大头腐蚀增加的趋势未达统计学显著。作者认为OxZr似乎能有效减少锥部损伤。

要点

  • 取回的OxZr股骨头平均腐蚀评分1.3 vs CoCr组2.1,差异显著(p<0.01)。
  • 取回的OxZr股骨头平均微动评分1.3 vs CoCr组1.5,差异显著(p=0.02)。
  • OxZr股骨头近端区域损伤较远端更重,而CoCr股骨头损伤与位置无关。

核心结论

  • 对取回OxZr股骨头的分析显示,其微动和腐蚀程度低于CoCr股骨头,OxZr似乎能有效减少锥部损伤。

研究设计与样本

  • 取回部件对比研究:28枚翻修全髋关节置换术中取回的OxZr股骨头,按患者人口学与28枚取回的CoCr股骨头匹配。
  • OxZr组置入时平均年龄56岁(46至83岁),CoCr组70岁(46至92岁)。
  • 采用改良Goldberg评分法测量股骨头female taper的微动与腐蚀评分。

主要发现

  • OxZr组取回股骨头平均腐蚀评分1.3(1至2.75)vs CoCr组2.1(1至4),p<0.01。
  • OxZr组平均微动评分1.3(1至2)vs CoCr组1.5(1至2.25),p=0.02。
  • OxZr股骨头近端区域损伤较股骨头远端区域更重,而位置对CoCr股骨头损伤无影响。
  • 仅在CoCr股骨头中观察到大头腐蚀增加的趋势,但未达统计学显著。

临床怎么用

  • 原文未据此给出推荐。

局限性

  • 原文未报告局限性。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

AIMS: The aim of this study was to evaluate fretting and corrosion in retrieved oxidized zirconium (OxZr; OXINIUM, Smith & Nephew, Memphis, Tennessee) femoral heads and compare the results with those from a matched cohort of cobalt-chromium (CoCr) femoral heads. PATIENTS AND METHODS: A total of 28 OxZr femoral heads were retrieved during revision total hip arthroplasty (THA) and matched to 28 retrieved CoCr heads according to patient demographics. The mean age at index was 56 years (46 to 83) in the OxZr group and 70 years (46 to 92) in the CoCr group. Fretting and corrosion scores of the female taper of the heads were measured according to the modified Goldberg scoring method. RESULTS: The OxZr-retrieved femoral heads showed significantly lower mean corrosion scores than the CoCr heads (1.3 (1 to 2.75) vs 2.1 (1 to 4); p < 0.01). Mean fretting scores were also significantly lower in the OxZr cohort when compared with the CoCr cohort (1.3 (1 to 2) vs 1.5 (1 to 2.25); p = 0.02). OxZr heads had more damage in the proximal region compared with the distal region of the head. Location had no impact on damage of CoCr heads. A trend towards increased corrosion in large heads was seen only in the CoCr heads, although this was not statistically significant. CONCLUSION: Retrieval analysis of OxZr femoral heads showed a decreased amount of fretting and corrosion compared with CoCr femoral heads. OxZr seems to be effective at reducing taper damage. Cite this article: Bone Joint J 2019;101-B:386-389.

来源:Bone Joint J,查看原文。版权归原作者所有。

引用

  1. Hampton C, Weitzler L, Baral E, et al. Do oxidized zirconium heads decrease tribocorrosion in total hip arthroplasty? A study of retrieved components. Bone Joint J. 2019. PMID: 30929492. doi:10.1302/0301-620X.101B4.BJJ-2018-1316.R1

供执业医师工作参考,不替代临床判断;数值与分级以指南原文为准。