文献速递关节外科

基于人口学变量的初次全膝关节置换假体尺寸计算器的前瞻性验证

Prospective Validation of a Demographically Based Primary Total Knee Arthroplasty Size Calculator.
PubMed 文献 · J Arthroplasty · 2019-03-07

前瞻性纳入474例择期初次全膝关节置换患者,用仅含性别、身高、体重的已发表算法预测股骨和胫骨假体尺寸。胫骨和股骨组件在±1号内准确率分别为87%和76%,±2号内分别为97%和95%。体重低于70 kg及股骨尺寸增量小于2.5 mm的植入系统是预测偏差超过2号的独立危险因素。

要点

  • 仅用性别、身高、体重预测全膝关节置换假体尺寸,胫骨±1号准确率87%,股骨±1号准确率76%。
  • 胫骨±2号准确率97%,股骨±2号准确率95%。
  • 体重低于70 kg(OR=2.47)和股骨尺寸增量小于2.5 mm的植入系统(OR=5.50)是预测偏差超过2号的危险因素。

核心结论

  • 该前瞻性系列验证了仅基于人口学变量的简单算法能以高准确率预测全膝关节置换的假体尺寸。
  • 作者认为外科医生可用该算法简化术前规划,减少不必要的托盘、试模和植入物库存,尤其在资源有限的社区或门诊环境。
  • 作者提出未来需进一步评估股骨尺寸差异小于2.5 mm的组件,以使该算法在全球更适用。

研究设计与样本

  • 前瞻性纳入474例连续接受择期初次全膝关节置换的患者。
  • 纳入4名外科医生,其中3名未参与此前的回顾性队列研究。
  • 将患者性别、身高和体重输入已发表的关节置换尺寸预测移动应用程序。
  • 将算法准确性与来自5种不同植入系统的实际植入股骨和胫骨组件尺寸进行比较。
  • 采用多变量回归分析识别模型预测不准确离群值的独立危险因素。

主要发现

  • 评估±1号内准确率时,胫骨组件准确率为87%(412/474),股骨组件准确率为76%(360/474)。
  • 评估±2号内准确率时,胫骨准确率为97%(461/474),股骨准确率为95%(450/474)。
  • 实际组件超出预测2号的风险因素包括体重低于70 kg(比值比=2.47,95%置信区间[1.21-5.06],P=0.01)。
  • 实际组件超出预测2号的风险因素还包括使用股骨尺寸增量小于2.5 mm的植入系统(比值比=5.50,95%置信区间[3.33-9.11],P<0.001)。

关键数值与推荐

  • 胫骨组件±1号内准确率87%(412/474),±2号内准确率97%(461/474)。
  • 股骨组件±1号内准确率76%(360/474),±2号内准确率95%(450/474)。
  • 体重低于70 kg的比值比=2.47,95%置信区间[1.21-5.06],P=0.01。
  • 股骨尺寸增量小于2.5 mm的植入系统的比值比=5.50,95%置信区间[3.33-9.11],P<0.001。

临床怎么用

  • 作者认为外科医生可用该算法简化术前规划流程,减少不必要的托盘、试模和植入物库存。
  • 作者认为该算法尤其在资源有限的社区或门诊环境中具有应用价值。
  • 作者提出未来需进一步评估股骨尺寸差异小于2.5 mm的组件,以使该算法在全球更适用。

局限性

  • 作者提出未来需进一步评估股骨尺寸差异小于2.5 mm的组件,以使该算法在全球更适用。

以上解读依据论文摘要(全文未开放获取或许可不允许翻译转载)。

原文

BACKGROUND: Preoperative planning for total knee arthroplasty (TKA) is essential for streamlining operating room efficiency and reducing costs. Digital templating and patient-specific instrumentation have shown some value in TKA but require additional costs and resources. The purpose of this study was to validate a previously published algorithm that uses only demographic variables to accurately predict TKA tibial and femoral component sizes. METHODS: Four hundred seventy-four consecutive patients undergoing elective primary TKA were prospectively enrolled. Four surgeons were included, three of which were unaffiliated with the retrospective cohort study. Patient sex, height, and weight were entered into our published Arthroplasty Size Prediction mobile application. Accuracy of the algorithm was compared with the actual sizes of the implanted femoral and tibial components from 5 different implant systems. Multivariate regression analysis was used to identify independent risk factors for inaccurate outliers for our model. RESULTS: When assessing accuracy to within ±1 size, the accuracies of tibial and femoral components were 87% (412/474) and 76% (360/474). When assessing accuracy to within ±2 sizes of predicted, the tibial accuracy was 97% (461/474), and the femoral accuracy was 95% (450/474). Risk factors for the actual components falling outside of 2 predicted sizes include weight less than 70 kg (odds ratio = 2.47, 95% confidence interval [1.21-5.06], P = .01) and use of an implant system with <2.5 mm incremental changes between femoral sizes (odds ratio = 5.50, 95% confidence interval [3.33-9.11], P < .001). CONCLUSIONS: This prospective series of patients validates a simple algorithm to predict component sizing for TKA with high accuracy based on demographic variables alone. Surgeons can use this algorithm to simplify the preoperative planning process by reducing unnecessary trays, trials, and implant storage, particularly in the community or outpatient setting where resources are limited. Further assessment of components with less than 2.5-mm differences between femoral sizes is required in the future to make this algorithm more applicable worldwide.

来源:J Arthroplasty,查看原文。版权归原作者所有。